Page 96 - Applied Pharmacognosy
P. 96

Types of Chemical markers

          HPLC Assay validation

•The assay is not valid unless:
1. the resolution factor between the peaks corresponding to

    aristolochicacid II (retention time about 6 minutes ) and
    aristolochicacid I (retention time about 7 minutes ) is at
    least 3.
2. Inject standard solution six times. The relative standard
    deviation of the areas of the peaks is at most 1.5%.

                                      Retention time (minutes)
   91   92   93   94   95   96   97   98   99   100   101