EOS          90398-A_PHY-0004-Frames and transformer assembly



               3.  Lateral side ladder frame

                   a.  Identify and install ladder frame behind the
                      lateral column. The ladder frame may be
                      partially pre-assembled prior to shipping.
                   b.  The frame at the back of the lateral
                      column is identifiable as it has an extra
                      bracket that holds the transformer at the

                      bottom.
                   c.  Both frames attach to the back of both
                      columns using 6X T-Nut bolts
                   d.  Once the frame is secured, install the
                      isolation transformer



                                The Isolation Transformer
                                weights about 70Kg. You
                                may need 2 people to
                                handle.





               4.  Install Isolation transformer

                   a.  The transformer mounts at the back of the
                      lateral column on the bottom shelf of the
                      metal frame. Use 4X (M10x20) bolts with

                      nuts and flat washers
                   b.  Again, the transformer weights about
                      70Kg, you will need a way to transport
                      and mount it.
                   c.  Make sure the transformer is mounted
                      properly so that it does not interfere with
                      the movement of the lift motor up/down.

                      Check the position of the mounting
                      brackets.
                   d.  Preferably, the primary power block on the transformer should be towards the outside (thick
                      cables 1U, 1V, 1W). Neutral only on the secondary side.
                   e.  Proceed to Lateral side
   43   44   45   46   47   48   49   50   51   52   53