Page 87 - Pricelist
P. 87

DETECTOR

            • FID or TCD
            • Temperature range: room temperature up
               7 °C – 420 °C
            • Max installation Qty: 2
            • Ignition model: automatic

            1. Hydrogen flame ionization detector (FID):
               ogarithmic amplifier
               High voltage switch control
               Baseline signal display
               Detection limit: 3 X 10  g/s
                                 -12
               (sample: n- hexadecane)
               Baseline noise: ≤ 5 X 10 -14  A
               Baseline draft: ≤6 X 10  A
                                  -13
               Dynamic range: 107
               RSD: ≤ 3%


            2. Thermal conductivity detector (TCD):
               Sensitivity: 8000mV.ml/mg (sample: n-hexadecane)
               Baseline noise: ≤ 0.05 mV
               Baseline draft: ≤ 0.15 mV/30min
               Dynamic range: 105

             OVERALL DIMENSION AND WEIGHT


            Power: 220V, 50Hz, 3000W
            Dimension: 1020 X 660 X 740 mm
            Net weight: 70 Kg











































                                             BioEra Life Sciences Pvt. Ltd.  |  www.bioera.in                 87