Page 9 - TEST_QUIRURGICAS_Neat
P. 9
Page 8 of 16 Ref. No.: PPE-20200015
EN 149: 2001 + A1: 2009
Clause Requirement – Test Result - Remark Verdict
The field of vision is acceptable if determined so in The vision does not be
P
practical performance tests. affected by the sample.
7.15 Exhalation valve(s) Without Exhalation valve N/A
A particle filtering half mask may have one or more
exhalation valve(s), which shall function correctly in all N/A
orientations.
If an exhalation valve is provided it shall be protected
against or be resistant to dirt and mechanical damage
N/A
and may be shrouded or may include any other device
that may be necessary for the particle filtering half mask.
Exhalation valve(s),if fitted, shall continue to operate
correctly after a continuous exhalation flow of 300 l/min N/A
over a period of 30 s.
When the exhalation valve housing is attached to the
face-blank, it shall withstand axially a tensile force of 10 N N/A
applied for 10 s.
7.16 Breathing resistance P
The breathing resistances apply to valved and valveless
particle filtering half masks and shall meet the P
requirements .
7.17 Clogging P
7.17.1 General N
For single-use devices only, the clogging test is an
N/A
optional test.
7.17.2 Breathing resistance P
7.17.2.1 Valved particle filtering half masks N/A
After clogging the inhalation resistances shall not exceed:
——FFP1: 4mbar
N/A
——FFP2: 5mbar
——FFP3: 7mbar
7.17.2.2 Valveless particle filtering half masks N/A
After clogging the inhalation and exhalation resistances
shall not exceed:
——FFP1: 3 mbar N/A
——FFP2: 4 mbar
——FFP3: 5 mbar
7.17.3 Filter penetration N/A
All types (valved and valveless) of particle filtering half
masks claimed to meet the clogging requirement shall
N/A
also meet the penetration requirements after the
treatment.