Page 12 - Catalogue - cuvettes and solid-state calibration standards
P. 12

Matching

 Means that the cuvettes are manufactured from materials that have transmission properties within 1% of each
     other. Accordingly, only the absorption directly through the lightpath of the cuvette is tested – other parameters
     used in fluorimetry are not addressed. At the point at which all parameters are sustained to a high degree the
     cuvette is “matched” by the fact that there is little distinction in any of the cells of the same physical
     configuration, lightpath and material. With more modern machines and stringently controlled manufacturing
     processes, Matching is not as important nowadays. FireflySci cells are manufactured within extremely tight
     tolerances that ensure superior matching.

Transmission

 Describes how well light passes through the optically polished windows of a cuvette at different wavelengths
     and is expressed in %. The measurement scale is 0-100%.

Z-dimension (Zd)

 The Z-dimension – also known as Zd or centre height – is the height from the very bottom of the cell to the
     centre of its sample chamber. Particularly if the cuvette’s polished window is very small (in thiscase, usually the
     entire cuvette is black except for the window) it’s critical that the Zd of the cuvette be the same as the height of
     the instrument’s light beam. Generally, Zd is either 8.5, 15 or 20 mm, depending on the instrument’s
     manufacturer.

 Only certain cuvettes depend on the correct Z-dimension and you’ll find the relevant information in their data
     tables. If you’re uncertain which Zd you need for your instrument, please ask us.