       Step-by-Step Instructions                                                                 Warnings                                                    Y-MK1-853


       6.4     Replacement of the Rechargeable Battery                                                   •  Inaccurate or incorrect readings due to the environment are likely to
                                                                                                            occur in the following cases:
               Propex Pixi™ is powered by one 1.2V AAA NiMH rechargeable                                    - Presence of portable or movable radio frequency transmitters in
               battery.
                                                                                                             the surroundings;
               If a fully charged battery is not sufficient for normal device operation                     - Film viewers or other Illumination devices which use an inverter
               during at least one working day, the battery should be replaced by a                          may cause abnormal operation of the apex locator. Such devices
               new one at your earliest convenience.                                                         should be turned off during use of the Propex Pixi™;
               Please note that a new battery should be charged for 24 hours prior to                       - Electromagnetic interference could cause improper operation of
               the first use of the device.                                                                  the device. In such cases the device behaviour may become
                                                                                                             abnormal or random. Usage of any devices emitting
               Warning                                                                                       electromagnetic radiation, such as cellular phones, remote
               Use GP100AAAHC or compatible NiMH rechargeable battery only.                                  controls, transceivers, etc., should be prohibited in the vicinity of
               Usage of non-rechargeable batteries may cause device damage.                                  Propex Pixi™.
               The battery compartment is located at the rear side of Propex Pixi™.                      •  General safety warnings:
               1. Carefully lift and tilt back the silicon cover to expose the screw.                       - In order to prevent infectious agent transfer it is highly
                  Release the screw.                                                                         recommended to use a rubber dam system during the endodontic
                                                                                                             procedure.
                                                                                                            - Make sure that the lip clip, hook or fork does not come into contact
                                                                                                             with an electric power source such as an electrical socket. This
                                                                                                             could result in a severe electrical shock.
                                                                                                            - Do not use Propex Pixi™ in the presence of flammable
                                                                                                             substances.
                                                                                                         •  Only use the original charger.
                                                                                                         •  Use AAA 1.2V 1000mAh NiMH rechargeable batteries only. Usage
                                                                                                            of non rechargeable batteries may cause device damage.
                                                                                                            Recommended battery type: Manufacturer – GP Batteries
                                                                                                                                      Model No.: GP100AAAHC.
                                            Fig. 5




















       14/222                 #        B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X)  25268-EYAL - 25268_pnim_eyal | 1 - A | 18-05-15 | 13:58:38 | SR:-- | Cyan   25268-EYAL - 25268_pnim_eyal | 1 - A | 18-05-15 | 13:58:38 | SR:-- | Yellow   25268-EYAL - 25268_pnim_eyal | 1 - A | 18-05-15 | 13:58:38 | SR:-- | Magenta   #25