 #25268-EYAL - 25268_pnim_eyal | 1 - A | 18-05-15 | 13:58:38 | SR:-- | Black 25268-EYAL - 25268_pnim_eyal | 1 - A | 18-05-15 | 13:58:38 | SR:-- | Yellow 25268-EYAL - 25268_pnim_eyal | 1 - A | 18-05-15 | 13:58:38 | SR:-- | Magenta 25268-EYAL - 25268_pnim_eyal | 1 - A | 18-05-15 | 13:58:38 | SR:-- | Cyan
 Y-MK1-853
 Troubleshooting  1  Step-by-Step Instructions

         6.6.3    Apical Zone
 #  Problem  Possible cause  Solution  2
                  The apical zone is divided into 3 bars graduated from 1.0 to 0.0 (apex)
 Perform the cable
 Bad electrical   connection test as   – see Fig. 11.
 contact.  described in the user
 manual, section 6.5.
 The connection   Place the connection hook
 hook is not properly   on the metal part of the file
 connected to the   below the plastic handle.
 file.
 The root canal is   Check the comparative
 obliterated.  X-ray image for hints.
 In the case of
 retreatment: old   Remove old root filling
 filling material   material residues prior to
 The transmission of the   residues may block   use.
 electric signal is   the root canal.  Beginning of apical zone  Middle of apical zone  Apical position
 interrupted. The device
 6  The root canal may
 does not show file                           Fig. 11
 progression inside the   be blocked by the   Completely remove the
 canal.  remnants of a   remnants prior to use.
 medication (e.g.   When the apex is reached, a solid tone is emitted.
 calcium hydroxide).  The indication 0.0 on the Propex Pixi™ screen relates to the minor
 Rinse the root canal with   apical foramen file position (the apical length).
 Root canal is   NaCl solution. Dry the
 extremely dry.  access cavity with a cotton
 pellet/ air-blower.  6.6.4  Over-Instrumentation
 If there is no parietal   A red “OVER” segment and an audio warning signal (rapid intermittent
 The selected file is   contact use larger ISO size   signal) indicate that the file has passed the apex – Fig. 12.
 too small for a large   file.
 root canal.  Important: exactly fitting
 files lead to precise results.
 Electronic   Contact your distributor.
 malfunction.











                                              Fig. 12


 34/222  B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X)  B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X)  19/222