Step-by-Step Instructions                                                                 Troubleshooting



       6.6.5   Completion of the measurements
                                                                                                   #          Problem           Possible cause          Solution
               •  Before unplugging the measurement cable from the device                                                                        Try pressing the ”ON/ OFF”
                  receptacle disconnect the lip clip and the connection hook from the                                         The button is      button several times. If the
                  patient.                                                                                                    functioning        device still does not turn
               •  Move the file stopper to the selected reference point on the tooth.                  The device does not turn   incorrectly.   on, please contact your
               •  Gently remove the file from the canal and measure the apical length              2   on by pressing the “ON /                  distributor.
                  between the stopper and the file.                                                    OFF” button.           The battery is
                                                                                                                              discharged.        Charge the battery.
               Note
               As a safety precaution in order to avoid over-instrumentation, it is                                           Electronic         Contact your distributor.
               recommended to proceed as follows: stop file progression in the                                                malfunction.
               canal at the point where the Propex Pixi™ indicates ‘0.0’.                          3   The device shuts off during   The battery is low.  Charge the battery.
               Place the file onto an endodontic ruler and measure the apical length.                  the procedure.
               Subtract a minimum of 0.5 mm from the measured file length.                             No sound during the    The sound control   Adjust the sound level by
                                                                                                   4                          is set at “Mute”   pressing the “VOLUME”
                                                                                                       procedure.             level.             button.
       6.7     Sound Adjustment
                                                                                                                              There is not a good   Ensure a good contact
               Propex Pixi™ is equipped with an audio indicator which enables                                                 contact between the   between the mucosa and
               monitoring of the progression of the file within the canal in addition to                                      lip clip and the oral   lip clip (Place the lip clip in
               visual monitoring.                                                                                             mucosa.            the labial angle opposite
                                                                                                                                                 the tooth to be treated).
               The volume can be adjusted to four different levels: mute, low, normal                                         The connection     Clean the connection hook
               and high, by successively pressing the    “VOLUME” button.                                                     hook is soiled.    (with Ethanol).
               When sound level is muted the    icon is turned off. At other sound
               levels the    icon remains turned on.                                                                          Deep caries
                                                                                                       Display is not steady   provides a        Block the external
                                                                                                   5                          conductive path    conductive path.
                                                                                                       during the procedure.
                                                                                                                              outside the canal.
                                                                                                                                                 Remove the file, close the
                                                                                                                                                 perforation and repeat the
                                                                                                                              Perforation.       apex detection procedure,
                                                                                                                                                 carefully inserting the file
                                                                                                                                                 into canal.
                                                                                                                                                 Try continuing the
                                                                                                                              Large lateral canal.  procedure by gently
                                                                                                                                                 advancing the file.











       20/222                          B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X)   B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X)           33/222                                     #                                                                25268-EYAL - 25268_pnim_eyal | 1 - B | 18-05-15 | 13:58:38 |