6.6.5 Completion of the measurements
# Problem Possible cause Solution
• Before unplugging the measurement cable from the device Try pressing the ”ON/ OFF”
receptacle disconnect the lip clip and the connection hook from the The button is button several times. If the
patient. functioning device still does not turn
• Move the file stopper to the selected reference point on the tooth. The device does not turn incorrectly. on, please contact your
• Gently remove the file from the canal and measure the apical length 2 on by pressing the “ON / distributor.
between the stopper and the file. OFF” button. The battery is
discharged. Charge the battery.
Note
As a safety precaution in order to avoid over-instrumentation, it is Electronic Contact your distributor.
recommended to proceed as follows: stop file progression in the malfunction.
canal at the point where the Propex Pixi™ indicates ‘0.0’. 3 The device shuts off during The battery is low. Charge the battery.
Place the file onto an endodontic ruler and measure the apical length. the procedure.
Subtract a minimum of 0.5 mm from the measured file length. No sound during the The sound control Adjust the sound level by
4 is set at “Mute” pressing the “VOLUME”
procedure. level. button.
6.7 Sound Adjustment
There is not a good Ensure a good contact
Propex Pixi™ is equipped with an audio indicator which enables contact between the between the mucosa and
monitoring of the progression of the file within the canal in addition to lip clip and the oral lip clip (Place the lip clip in
visual monitoring. mucosa. the labial angle opposite
the tooth to be treated).
The volume can be adjusted to four different levels: mute, low, normal The connection Clean the connection hook
and high, by successively pressing the “VOLUME” button. hook is soiled. (with Ethanol).
When sound level is muted the icon is turned off. At other sound
levels the icon remains turned on. Deep caries
Display is not steady provides a Block the external
5 conductive path conductive path.
during the procedure.
outside the canal.
Remove the file, close the
perforation and repeat the
Perforation. apex detection procedure,
carefully inserting the file
into canal.
Try continuing the
Large lateral canal. procedure by gently
advancing the file.
20/222 B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X) B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X) 33/222 # 25268-EYAL - 25268_pnim_eyal | 1 - B | 18-05-15 | 13:58:38 |