#25268-EYAL - 25268_pnim_eyal | 1 - B | 18-05-15 | 13:58:38 | SR:-- | Black 25268-EYAL - 25268_pnim_eyal | 1 - B | 18-05-15 | 13:58:38 | SR:-- | Yellow 25268-EYAL - 25268_pnim_eyal | 1 - B | 18-05-15 | 13:58:38 | SR:-- | Magenta 25268-EYAL - 25268_pnim_eyal | 1 - B | 18-05-15 | 13:58:38 | SR:-- | Cyan
Cleaning, Disinfection and Sterilization Cleaning, Disinfection and Sterilization
Operation Operating mode Warning Operation Operating mode Warning
• Once the accessories • The hook mechanism has to • Pulp and dentine • Do not allow remnants to dry
have been cleaned be activated several times remnants must be on. Clean no later than 2
and inspected, place during disinfection and the removed from the hours. For visible impurities
them into the rinse to allow the inner parts accessories that are observed on
disinfection bath for to be disinfected more immediately. instruments, a pre- cleaning
the prescribed effectively. • After using the is recommended by brushing
contact time (e.g. • To dry use oil-free, filtered accessories on the them manually with soft
Cidex OPA, Johnson compressed air and then patient, place them material. Only use clean, soft
& Johnson Medical, leave the accessories to dry directly into a dish brushes to manually remove
100% for 20 minutes); further in a clean place for at filled with a suitable contamination and remnants,
the accessories must least 20 minutes. cleaning and or a clean, soft cloth or wipe
be sufficiently • Once the accessories are disinfection solution that is only used for this
5 Disinfection
®
immersed in the dry, inspect and pack them (e.g. CIDEZYME , purpose. Do not use metal
solution. as soon as possible. ENZOL Enzymatic brushes or wire wool. Check
®
• Remove the Detergent Solutions, that no visible contamination
accessories from the Johnson & Johnson or remnants remain, and
disinfection bath and Medical, 0.8% for repeat the pre-cleaning
rinse them off 2 Pre-Disinfection between 1 minute process if necessary.
thoroughly at least and 2 hours) for • Make sure that the products
five times for a minute cleaning, pre- are fully immersed.
each time with water. disinfection and • The disinfectant solution
• Dry the accessories interim storage. should be aldehyde free (to
by blowing them • Wash the avoid blood impurities
down fully. fixation), suitable for
accessories under
• Abundant rinsing • Use quality water in flowing sterile, disinfecting the accessories,
(at least 1 min). accordance with local deionized water or in and compatible with the
regulations. a disinfection solution accessories.
• If a disinfecting solution at least three times • Disinfectant used for pre-
contains a corrosion for a minute each treatment is only for personal
inhibitor, it is recommended time, in order to protection and does not
6 Rinsing obviate the need for
to rinse the instruments just remove all visible
before the autoclaving. traces of disinfection once cleaning
• Dry on a single use non- contamination and has been completed.
weaved cloth, or with a remnants. Pre-treatment should never
drying machine or filtered be omitted.
compressed air. • Abundant rinsing • Use clean water.
(at least 1 min). • If a pre-disinfectant solution
contains a corrosion
3 Rinsing
inhibitor, it is recommended
to rinse the instruments just
before the cleaning.