 Cleaning, Disinfection and Sterilization  Cleaning, Disinfection and Sterilization



         7        Cleaning, Disinfection and Sterilization
 Operation  Operating mode  Warning
 •  Steam sterilization at:   •  The accessories (lip clip,   7.1  General Recommendations
 134 °C / 273°F,   hook and fork) must be
 3 min.  sterilized according to the   •  The device does not contain user serviceable parts. The service
 packaging labelling.  and repair should be provided by factory trained service personnel
 •  Use only autoclaves that   only.
 match the requirements of   •  After each use, all objects that were in contact with infectious
 EN 13060, EN 285.
 •  Respect the maintenance   agents should be cleaned using towels impregnated with a
 procedure of the autoclave   disinfecting and detergent solution (a bactericidal, fungicidal and
 device given by the   aldehyde free solution). Use of chemical agents may cause
 9  Sterilization
 manufacturer.       damage to the equipment. We recommend to use only a
 •  Use only this recommended   disinfecting solution which is approved for its efficacy
 sterilization procedure.  (VAH/DGHM-listing, CE marking, FDA approval).
 •  Control the efficiency   •  The lip clip and the hook must be sterilized between treatments.
 (packaging integrity, no   Please note that the measuring cable cannot be autoclaved.
 humidity, colour change of   •  In addition, a fork is not included in the Propex Pixi™ but it can be
 sterilization indicators,
 physico-chemical    used and should follow the same procedure as the lip clip and hook.
 integrators, digital records of   •  Follow the “Disinfection and sterilization procedure” described in
 cycles parameters).  section 7.2.
 •  Keep devices in   •  Sterility cannot be   •  The user is responsible for the sterility of the lip clip, the connection
 sterilization   guaranteed if packaging is   hook and the fork for the first cycle and each further usage.
 packaging in a dry   open, damaged or wet.  •  All damaged accessories should be discarded and dirty
 10  Storage  and clean   •  Check the packaging and the   accessories should be cleaned and sterilized per the procedure
 environment.  medical devices before using   described in section 7.2.
 them (packaging integrity, no
 humidity and validity period).























 30/222  #  B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X)  25268-EYAL - 25268_pnim_eyal | 1 - A | 18-05-15 | 13:58:38 | SR:-- | Yellow   25268-EYAL - 25268_pnim_eyal | 1 - A | 18-05-15 | 13:58:38 | SR:-- | Cyan   25268-EYAL - 25268_pnim_eyal | 1 - A | 18-05-15 | 13:58:38 | SR:-- | Magenta   #25268-EYAL - 25268_pnim_eyal