Page 207 - 25268
P. 207

故障排除                                                         Y-MK1-853




 Recommended separation distances between portable and mobile RF   #  故障  可能的原因  解决措施
 communications equipment and the not life support equipment  按照 《用户手册》中章节
                                      电气接触不良。
 The Product is intended for use in an electromagnetic environment in which radiated   6.5 所述进行线缆连接测试。
 RF disturbances are controlled. The customer or the user of the Product can help   连接钩未妥当连接至  将连接钩置于塑料手柄下挫
 prevent electromagnet interference by maintaining a minimum distance between   挫针。  针的金属部分。
 portable and mobile RF communications equipment (transmitters) and the Product as   根管闭塞。  对照 X 线图像,查找线索。
 recommended below, according to the maximum output power of the communications
 equipment.                           如果是再次治疗:旧          使用前先清除旧的填充材料
                                      的填充材料残留物可
 Separation distance according to frequency of transmitter   能堵塞根管。  残留物。
 m             电子信号传输受干扰。设备
 Rated maximum   6  不显示挫针在根管内的前       根管可能被药物残留          使用前彻底清除残留物。
 output power of   150 kHz to  80 MHz to  800 MHz to  进。  堵塞 ( 如氢氧化钙 )。
 transmitter W  80 MHz  800 MHz  2.5 GHz                 使用 NaCl 溶液冲洗根管。
 d =  0.35 ×  P  d =  0.35 ×  P  d =  0.7 ×  P  根管极其干燥。  用棉球 / 吹风机使开髓口干
                                                         燥。
 0.01  0.04  0.04  0.07                                  如果没有腔壁接触,则可使
 0.1  0.11  0.11  0.22                牙根较大,而所选锉          用较大的 ISO 尺寸挫针。
                                      针过小。               切记:挫针精准安放,才能
 1  0.35  0.35  0.7
                                                         获得精准的结果。
 10  1.1  1.1  2.2                    电子失灵。              联系您的供销商。
 100  3.5  3.5  7
 For transmitters rated at a maximum output power not listed above, the recommended
 separation distance d in metres (m) can be estimated using the equation applicable to
 the frequency of the transmitter, where P is the maximum output power rating of the
 transmitter in watts (W) according to the higher frequency range applies.
 Note
 At 80 MHz and 800 MHz, the separation distance fort the
 higher frequency range applies.
 Note
 These guidelines may not apply in all situations.
 Electromagnetic propagation is affected by absorption
 and reflection from structures, objects and people.















 220/222  #  B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X)  25268-EYAL - 25268_pnim_eyal | 7 - B | 18-05-15 | 13:58:39 | SR:-- | Magenta   25268-EYAL - 25268_pnim_eyal | 7 - B | 18-05-15 | 13:58:39 | SR:-- | Yellow   25268-EYAL - 25268_pnim_eyal | 7 - B | 18-05-15 | 13:58:39 | SR:-- | Cyan   #25268-EYAL - 25268_pnim_eyal