 #25268-EYAL - 25268_pnim_eyal | 7 - B | 18-05-15 | 13:58:39 | SR:-- | Black 25268-EYAL - 25268_pnim_eyal | 7 - B | 18-05-15 | 13:58:39 | SR:-- | Yellow 25268-EYAL - 25268_pnim_eyal | 7 - B | 18-05-15 | 13:58:39 | SR:-- | Magenta 25268-EYAL - 25268_pnim_eyal | 7 - B | 18-05-15 | 13:58:39 | SR:-- | Cyan



         Appendix
 Guidance and manufacturer’s declaration - electromagnetic emissions
 The Product is intended for use in the electromagnetic environment specified below.   ELECTROMAGNETIC COMPATIBILITY (EMC)
 The customer or the user of the Product should assure that it is used in such an
 environment.

 Electromagnetic   Changes or modifications to this product not expressly approved by the
 IEC 60601
 Immunity tests  Compliance level  environment -  manufacturer may result in increased emissions or decreased immunity
 Test level
 guidance         performance of the product and could cause EMC issues with this or
                  other equipment. This product is designed and tested to comply with
 Floors should be
 Electrostatic    applicable regulations regarding EMC and shall be installed and put into
 discharge (ESD)  wood, concrete or   service according to the EMC information stated as follows.
 ceramic tile. If
 +/- 6 kV contact  +/- 6 kV contact  floors are covered
 with synthetic
 +/- 8 kV air  +/- 8 kV air  material, the   Warning
 relative humidity   Use of portable phones or other radio frequency (RF) emitting
 should be at least   equipment near the product may cause unexpected or adverse
 IEC 61000-4-2
 30 %.            operation.
 Power frequency   Warning
 magnetic fields   The product must not be used adjacent to, or stacked with, other
 Power frequency   should be at levels   equipment. If adjacent or stacked use is necessary, the product must
 (50/60Hz)   3 A/m  3 A/m  characteristic of a   be tested to verify normal operation in the configuration in which it is
 magnetic field  typical location in a
 IEC 61000-4-8  typical commercial   being used.
 or hospital
 environment.
 Note
 UT is the a.c. mains voltage prior to application of the test
 level.