Mains power quality Appendix 1
25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - B | 18-06-11 | 09:03:44 | SR:-- | Magenta
should be that of a typical
25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - B | 18-06-11 | 09:03:44 | SR:-- | Yellow
#25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - B | 18-06-11 | 09:03:44 | SR:-- | Black
25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - B | 18-06-11 | 09:03:44 | SR:-- | Cyan
Voltage dips 0% U ; 0.5 cycle 0% U ; 0.5 cycle public low-voltage power ELECTROMAGNETIC COMPATIBILITY
T
T
supply network that
EMC according to IEC 60601-1-2 ed.3
supplies buildings used
for domestic purposes,
0% U ; 1 cycle 0% U ; 1 cycle commercial or hospital, Changes or modifications to this product not expressly approved by the
T
T
and and
Voltage clinical environment. If manufacturer may result in increased emissions or decreased immunity
T
T
interruptions 70% U ; 25/30 70% U ; 25/30 the user of the performance of the product and could cause EMC issues with this or
cycles cycles ProPex II requires other equipment. This product is designed and tested to comply with
®
battery charging during applicable regulations regarding EMC and shall be installed and put into
power mains service according to the EMC information stated as follows.
interruptions it is
0% U ; 250/300 0% U ; 250/300
IEC T T recommended that the
61000-4-11 cycles cycles ProPex II charger be
®
powered from a separate Use of portable phones or other radio frequency (RF) emitting
power supply (UPS, etc.). equipment near the product may cause unexpected or adverse
operation.
Power frequency
magnetic fields should be The product must not be used adjacent to, or stacked with, other
at levels characteristic of equipment. If adjacent or stacked use is necessary, the product must
Rated power a typical public low- be tested to verify normal operation in the configuration in which it is
frequency 30 A/m 30 A/m voltage power supply being used.
magnetic 50 or 60 Hz 50 or 60 Hz network that supplies
fields buildings used of
domestic purposes,
commercial or hospital,
clinic environment.
NOTE: U is the a.c. mains voltage prior to application of the test level.
T
180/184 B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X) B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X) 169/184