Page 170 - fif
P. 170

#25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - A | 18-06-11 | 09:03:44 | SR:-- | Black
                                                                                                                                                                              25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - A | 18-06-11 | 09:03:44 | SR:-- | Yellow
                                                                                                                                                                              25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - A | 18-06-11 | 09:03:44 | SR:-- | Magenta
                                                                                                                                                                              #

               Compliant Cables and Accessories                                                                                                                               25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 7 - A | 18-06-11 | 09:03:44 | SR:-- | Cyan

               Use of non-original cables or accessories may result in increased                         Guidance and Manufacturer's declaration - Electromagnetic Immunity
               emissions or decreased immunity performance of the product.
                                                                                                             ®
                                                                                                  The ProPex  II is intended for use in the electromagnetic environment specified
                                                                                                                                        ®
                                                                                                  below. The customer or the user of the ProPex  II should assure that it is used in
               The table below lists cables and accessories for which the manufacturer            such an environment.
               claims EMC compliance:                                                             Immunity test  IEC 60601-1-2   Compliance level     Electromagnetic
                                                                                                                   Test level                      environment - guidance
            Description                          Details                                                                                          Floors should be wood,
                                                                                                  Electrostatic                                   concrete or ceramic tile.
        Measurement cable   Original only.                                                        discharge    ± 8 kV contact    ± 8 kV contact   If floors are covered with
        Accessories:                                                                              (ESD) IEC    ± 15 kV air       ± 15 kV air      synthetic material, the
                                                                                                  61000-4-2                                       relative humidity should
        Lip Clip            Original only.
                                                                                                                                                  be at least 30%.
        Connection Hook     Original only.
                                                                                                                                                  Mains power quality
        Connection Fork     Original only.                                                                     ± 2 kV for power                   should be that of a typical
                            Original Switching charger only:                                     Electrical fast   supply lines  ± 2 kV for power   public low-voltage power
        Charger             Input: 100 - 240 VAC / ~ 50-60 Hz                                     transients /                   supply lines     supply network that
                            Output: 6 VDC ± 5%, 1000 mA                                           bursts IEC   ± 1 kV for input/                  supplies buildings used
                                                                                                  61000-4-4                      Not applicable   for domestic purposes,
                                                                                                               output lines
                                                                                                                                                  commercial or hospital,
                                                                                                                                                  clinic environment
                                                                                                                                                  Mains power quality
                                                                                                                                                  should be that of a typical
                                                                                                               ± 1 kV Line-to-line  ± 1 kV Line-to-line  public low-voltage power
                                                                                                  Surges IEC                                      supply network that
                                                                                                  61000-4-5    ± 2 kV Line-to-   ± 2 kV Line-to-  supplies buildings used
                                                                                                               ground            ground           for domestic purposes,
                                                                                                                                                  commercial or hospital,
                                                                                                                                                  clinic environment


















       170/184                          B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X)  B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X)           179/184
   165   166   167   168   169   170   171   172   173   174   175