Use of non-original cables or accessories may result in increased Guidance and Manufacturer's declaration - Electromagnetic Immunity
emissions or decreased immunity performance of the product.
®
The ProPex II is intended for use in the electromagnetic environment specified
®
below. The customer or the user of the ProPex II should assure that it is used in
The table below lists cables and accessories for which the manufacturer such an environment.
claims EMC compliance: Immunity test IEC 60601-1-2 Compliance level Electromagnetic
Test level environment - guidance
Description Details Floors should be wood,
Electrostatic concrete or ceramic tile.
Measurement cable Original only. discharge ± 8 kV contact ± 8 kV contact If floors are covered with
Accessories: (ESD) IEC ± 15 kV air ± 15 kV air synthetic material, the
61000-4-2 relative humidity should
Lip Clip Original only.
be at least 30%.
Connection Hook Original only.
Mains power quality
Connection Fork Original only. ± 2 kV for power should be that of a typical
Original Switching charger only: Electrical fast supply lines ± 2 kV for power public low-voltage power
Charger Input: 100 - 240 VAC / ~ 50-60 Hz transients / supply lines supply network that
Output: 6 VDC ± 5%, 1000 mA bursts IEC ± 1 kV for input/ supplies buildings used
61000-4-4 Not applicable for domestic purposes,
output lines
commercial or hospital,
clinic environment
Mains power quality
should be that of a typical
± 1 kV Line-to-line ± 1 kV Line-to-line public low-voltage power
Surges IEC supply network that
61000-4-5 ± 2 kV Line-to- ± 2 kV Line-to- supplies buildings used
ground ground for domestic purposes,
commercial or hospital,
clinic environment
170/184 B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X) B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X) 179/184