Cleaning, Disinfection and sterilization Cleaning, Disinfection and sterilization
7.2 General Recommendations
N° Operation Operating mode Warning
1. Once the accessories have • The user is responsible for the sterility of the product for the first
been cleaned and inspected, The hook mechanism has to be cycle and each further usage as well as for the usage of damaged
place them into the disinfection activated several times during
bath for the prescribed contact disinfection and the rinse to or dirty instruments.
time (e.g. Cidex OPA, Johnson allow the inner parts to be • For your own safety, please wear personal protective equipment
& Johnson Medical, 100% for disinfected more effectively.
20 minutes); the accessories (gloves, glasses, mask).
must be sufficiently immersed in To dry use oil-free, filtered • Use only disinfecting solution which is approved for their efficacy
4. Disinfection the solution. compressed air and then leave (VAH/DGHM-listing, CE marking, FDA approval).
the accessories to dry further in
2. Remove the accessories from a clean place for at least
the disinfection bath and rinse 20 minutes. The cleaning, disinfection and sterilization process applies only to the
them off thoroughly at least five lip clip, the hook and the fork. As long as they are treated with due care
times for a minute each time with Once the accessories are dry,
water. and are not damaged or contaminated, these accessories can be
inspect and pack them as soon
as possible. reused multiple times.
3. Dry the accessories by
blowing them down fully.
Make sure undamaged sterile packaging is used.
Use quality water in accordance
with local regulations.
If a disinfecting solution contains
a corrosion inhibitor, it is
5. Rinsing Abundant rinsing (at least recommended to rinse the
1 min). instruments just before the
autoclaving.
Dry on a single use non- weaved
cloth, or with a drying machine or
filtered compressed air.
Defects include:
- deformation of the plastic;
- corrosion;
- discoloration of the plastic.
6. Inspection Inspect devices and sort out
those with defects. Dirty instruments must be
cleaned and disinfected again.
Maintenance is not required.
Do not use instrument lubricant.
28/184 B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X) B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X) 25/184