 Cleaning, Disinfection and sterilization  Cleaning, Disinfection and sterilization
         7.2      General Recommendations
 N°  Operation  Operating mode  Warning
 1. Once the accessories have   •  The user is responsible for the sterility of the product for the first
 been cleaned and inspected,   The hook mechanism has to be   cycle and each further usage as well as for the usage of damaged
 place them into the disinfection   activated several times during
 bath for the prescribed contact   disinfection and the rinse to   or dirty instruments.
 time (e.g. Cidex OPA, Johnson   allow the inner parts to be   •  For your own safety, please wear personal protective equipment
 & Johnson Medical, 100% for   disinfected more effectively.
 20 minutes); the accessories   (gloves, glasses, mask).
 must be sufficiently immersed in   To dry use oil-free, filtered   •  Use only disinfecting solution which is approved for their efficacy
 4.  Disinfection  the solution.  compressed air and then leave   (VAH/DGHM-listing, CE marking, FDA approval).
 the accessories to dry further in
 2. Remove the accessories from   a clean place for at least
 the disinfection bath and rinse   20 minutes.  The cleaning, disinfection and sterilization process applies only to the
 them off thoroughly at least five   lip clip, the hook and the fork. As long as they are treated with due care
 times for a minute each time with   Once the accessories are dry,
 water.           and are not damaged or contaminated, these accessories can be
 inspect and pack them as soon
 as possible.     reused multiple times.
 3. Dry the accessories by
 blowing them down fully.
                  Make sure undamaged sterile packaging is used.
 Use quality water in accordance
 with local regulations.
 If a disinfecting solution contains
 a corrosion inhibitor, it is
 5.  Rinsing  Abundant rinsing (at least   recommended to rinse the
 1 min).  instruments just before the
 autoclaving.
 Dry on a single use non- weaved
 cloth, or with a drying machine or
 filtered compressed air.
 Defects include:
 - deformation of the plastic;
 - corrosion;
 - discoloration of the plastic.
 6.  Inspection  Inspect devices and sort out
 those with defects.  Dirty instruments must be
 cleaned and disinfected again.
 Maintenance is not required.
 Do not use instrument lubricant.