       Cleaning, Disinfection and sterilization                                                  Cleaning, Disinfection and sterilization
                                                                                                 7.2     General Recommendations
        N°        Operation          Operating mode            Warning
                                 1. Once the accessories have                                            •  The user is responsible for the sterility of the product for the first
                                 been cleaned and inspected,   The hook mechanism has to be                 cycle and each further usage as well as for the usage of damaged
                                 place them into the disinfection   activated several times during
                                 bath for the prescribed contact   disinfection and the rinse to            or dirty instruments.
                                 time (e.g. Cidex OPA, Johnson   allow the inner parts to be             •  For your own safety, please wear personal protective equipment
                                 & Johnson Medical, 100% for   disinfected more effectively.
                                 20 minutes); the accessories                                               (gloves, glasses, mask).
                                 must be sufficiently immersed in   To dry use oil-free, filtered        •  Use only disinfecting solution which is approved for their efficacy
        4.    Disinfection       the solution.          compressed air and then leave                       (VAH/DGHM-listing, CE marking, FDA approval).
                                                        the accessories to dry further in
                                 2. Remove the accessories from   a clean place for at least
                                 the disinfection bath and rinse   20 minutes.                           The cleaning, disinfection and sterilization process applies only to the
                                 them off thoroughly at least five                                       lip clip, the hook and the fork. As long as they are treated with due care
                                 times for a minute each time with   Once the accessories are dry,
                                 water.                                                                  and are not damaged or contaminated, these accessories can be
                                                        inspect and pack them as soon
                                                        as possible.                                     reused multiple times.
                                 3. Dry the accessories by
                                 blowing them down fully.
                                                                                                         Make sure undamaged sterile packaging is used.
                                                        Use quality water in accordance
                                                        with local regulations.
                                                        If a disinfecting solution contains
                                                        a corrosion inhibitor, it is
        5.    Rinsing            Abundant rinsing (at least   recommended to rinse the
                                 1 min).                instruments just before the
                                                        autoclaving.
                                                        Dry on a single use non- weaved
                                                        cloth, or with a drying machine or
                                                        filtered compressed air.
                                                        Defects include:
                                                        - deformation of the plastic;
                                                        - corrosion;
                                                        - discoloration of the plastic.
        6.    Inspection         Inspect devices and sort out
                                 those with defects.    Dirty instruments must be
                                                        cleaned and disinfected again.
                                                        Maintenance is not required.
                                                        Do not use instrument lubricant.



















       28/184                           B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X)  B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X)           25/184