Step-by-Step Instructions  Troubleshooting                                             #



 6.6.3  Over-Instrumentation
           N°        Problem            Possible cause           Solution
 A red segment and a warning signal indicate that the file has passed the   No sound during the   The sound control is set at   Adjust the sound level as
 Apex. The cursor on the tooth icon shows “OVER”.  5.  procedure.  “Mute” level.  described in the User Manual.
                                                           Ensure a good contact between
                                    There is not a good contact   the mucosa and lip clip (Place
                                    between the lip clip and the oral   the lip clip in the labial angle
                                    mucosa.
                                                           opposite the tooth to be treated).
                                                           Clean the connection hook (with
                                    The connection hook is soiled.
                                                           Ethanol).
                Indication of file position is   Deep caries provides a
           6.   not steady during the   conductive path outside the   Block the external conductive
                procedure.          canal.                 path.
                                                           Remove the file, close the
                                                           perforation and repeat the apex
                                    Perforation.
                                                           detection procedure, carefully
                                                           inserting the file into canal.
                                                           Try continuing the procedure by
                                    Large lateral canal.
                                                           gently advancing the file.
                                                           Perform the cable connection
                                    Bad electrical contact.  test as described in the User
                                                           Manual.
                                                           Place the connection hook on
                                    The connection hook is not   the metal part of the file below
                                    properly connected to the file.
 6.6.4  Completion of the Measurements                     the plastic handle.          25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 1 - B | 18-06-11 | 09:03:43 | SR:-- | Magenta
                                                           Check the comparative X-ray  25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 1 - B | 18-06-11 | 09:03:43 | SR:-- | Yellow  25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 1 - B | 18-06-11 | 09:03:43 | SR:-- | Cyan
 Before unplugging the Measurement cable from the device receptacle,   The root canal is obliterated.  image for hints.  #25567_eyal-Y-MK1-837 - 25567_pnim_y-mk1-837_eyal.1.1 | 1 - B | 18-06-11 | 09:03:43 | SR:-- | Black
 disconnect the lip clip and the connection hook from the patient.  In the case of re-treatment: old
                                    filling material residues may   Remove old root filling material
 Move the file stopper to the selected reference point on the tooth.  The device does not show   block the root canal.  residues prior to use.
           7.   file progression inside the
 Gently remove the file from the canal and measure the apical length   canal.  The root canal may be blocked   Completely remove the
 between the stopper and the file tip.  by the remnants of a medication   remnants prior to use.
                                    (e.g. calcium hydroxide).
 As a safety precaution in order to avoid over-instrumentation, it is   Root canal is extremely dry.  Rinse the root canal with NaCl
                                                           solution. Dry the access cavity
 recommended to proceed as follows:                       with a cotton pellet/ air-blower.
  ®
 - stop file progression in the canal at the point where the ProPex  II   If there is no parietal contact use
 indicates ‘APEX’;                 The selected file is too small for   larger ISO size file.
                                                           Important: exactly fitting files
                                    a large root canal.
 - place the file onto an endodontic ruler and measure the apical   lead to precise results.
 length;                           Electronic malfunction.  Contact your distributor.
 - subtract a minimum of 0.5 mm from the measured file length.











 20/184  B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X)  B EU PIIS DFU PRT / Rev.04 / 02-2018 (Old ZF 1902127.X)  33/184