Page 139 - Rapid Review of ECG Interpretation in Small Animal Practice, 2nd Edition
P. 139

Answer 57                   New Advanced-Level ECG Cases



           Answer 57
  VetBooks.ir   1  ECG 57a shows that the pacemaker is oversensing. The pacemaker occasionally detects the T wave

             and mistakes it for an intrinsically generated QRS complex. The pacemaker is programmed to pace at
             75 bpm (i.e., time between ventricular paced [VP] beats of 800 ms). In this example, the pacemaker
             senses the T wave (VS = ventricular sense, horizontal blue arrow) and waits to pace again 800 ms after
             the sensed T wave (ECG 57b). This results in a R-R interval that is longer than 800 ms (green arrow)
             and heart rate slower than 75 bpm each time the pacemaker senses the T waves.
            2  Oversensing can result in pacing at a lower than expected rate as the inappropriate detection inhibits
             the pacemaker from pacing at the correct rate. As described in Question 56, sensitivity refers to the
             minimum electrical voltage that the pacemaker will detect as a native depolarization. Oversensing
             occurs if the pacemaker senses electric signals, such as skeletal muscle myopotentials, electromagnetic
             interference, or P or T waves and interprets them as native QRS complexes. The likelihood of
             oversensing can be increased by inadequate programming of sensitivity settings or refractory periods,
             lead or device failure, or changes in signal amplitude due to change in tissue characteristics that can
             occur over time (i.e., fibrosis. hyperkalemia). In most cases, the inappropriately sensed electrical
             events are of lower voltage than the true QRS complexes, and decreasing the pacemaker’s ventricular
             sensitivity will correct the problem. In this case, the sensing threshold value was increased from 3.8 to
             5.0 mV, and the oversensing was corrected.



       57b
                 I



                                800 ms    VS                        VS
                 II
                                              800 ms


                        VP        VP            VP         VP             VP
                 III




                 aVR





                 aVL




                 aVF





            v:               Speed: 25 mm/sec    Limb: 10 mm/mV   Chest: 10.0 mm/mV         F  60~  0.





      126
   134   135   136   137   138   139   140   141   142   143   144