Page 18 - PBL71 OI/AXM630-EN AZTEC 600 SS (WIRO)
P. 18

Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers 3 Installation
3.5.1 Connections Overview
                                           AC Power Supply Earth (Ground) Stud (on AC models only)
– see Warnings in Section 3.5, page 13
                                               Sockets for Optional Digital Communications
                                                                                                                         TB2
           Ethernet RJ45 Connector Socket
Attention Relay
Failure Relay
NO NOC CNC NC
NO NOC
CNC NC
Stop Relay
Calibration in Progress Relay
Alarm 1 Relay
Alarm 3 Relay
Alarm 5 Relay
**Fuse 12.5 A Type T
TB1
          TB3
                –
O/P6
+
–
                TB4
          O/P5 NO
 +
–O/P4 +
–O / P 3 +
–O/P2 +
–O / P 1 +
TB8
Optional Digital Communications
Use fuse rating:
AC Supply: 1 A (max.), Type T, 250 V *DC Supply: 12.5 A 125 V DC Type T
*Manufacturer: SCHURTER, Model SPT 5 x 20 Series
NO
C
        TB7
Current Outputs
Alarm 2
Alarm 4
Alarm 6
Relay
Relay
Relay
CNC NC
NO NOC
CNC NC
NO NO C
CNC NC
             TB5
                                TB6
                         Fuse 1 A Type T
OR
       L N
+ –
E
    E 100 V to 240 V AC ±10 %
(90 V min. to 264 V max.) 50 / 60 Hz
18 to 36 V DC see Section 13.5, page 69, for DC fuse replacement details
  **
 Fig. 3.5 Connections Overview
14 OI/AXM630–EN Rev. J


































   16   17   18   19   20