Page 33 - PBL71 OI/AXM630-EN AZTEC 600 SS (WIRO)
P. 33
Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers 6 Configuration
6.1.3 Time
Fields
Description
Date and Time
Warning. Changing the time can result in the permanent loss of data. Once it is changed a warning is displayed stating that recording is disabled until the configuration has been saved.
Daylight Saving – Enable
Enables automatic daylight saving time adjustment. Options are:
Off. The 'Daylight Saving – Start' and 'Daylight Saving – End' fields are not available.
Auto – USA. The start and end of the daylight saving period in the USA is calculated automatically. The clock is incremented automatically by 1 hour at 2:00 am on the second Sunday in March and decremented automatically by 1 hour at 2:00 am on the first Sunday in November.
Auto – Europe. The start and end of the daylight saving period in Central Europe is calculated automatically. The clock is incremented automatically by 1 hour at 2:00 am on the last Sunday in March and decremented automatically by 1 hour at 2:00 am on the last Sunday in October.
Auto – Custom. The start and end date and time can be edited.
Daylight Saving – Start
If 'Daylight Saving – Enable' is set to USA or Europe, the start date is displayed but cannot be edited.
If 'Daylight Saving – Enable' is set to 'Custom' the date and time can be edited.
Daylight Saving – End
If 'Daylight Saving – Enable' is set to USA or Europe, the end date is displayed but cannot be edited.
If 'Daylight Saving – Enable' is set to 'Custom' the date and time can be edited.
OI/AXM630–EN Rev. J 29