CALIBRATION REPORT CERT NO. : DC19/0831
PAGE: 3 OF 3
RFSTT.T OF' CATTBRATTON : ADruSTMENT ( NO )
FTJNCTION: OUTSIDE MEASUREMENT
SCALE RANGE : 0 mm TO 300
RESOLUTION: 0.05 fllm
STANDARD LENGTII IruC READING ERROR
(tnm) ( ttu" ) (mm)
ZERO 0.00 0.00
0.5 0.50 0.00
1.4 1.40 0.00
5.1 5.10 0.oo
7.7 7.70 0.00
10.3 10.30 0.00
22.9 22.90 0.00
37.6 37.60 0.00
70.2 70.20 0.00
122.8 L22.80 0.00
200 200.00 o.o0
300 300.00 0.00
FTJNCTION : INSIDE MEASUREMENT
STANDARD LENGTH I.ruC READING ERROR
(tnn) ( rnrn ) (to-)
15.5 75.45 -0.05
16.4 16.35 -0.05
20.1 20.05 -0.05
22,7 22.65 -0.05
30.3 30.25 -0.05
37.9 37.85 -0.05
67.6 67.55 -0.05
95.2 95.15 -0.o5
r72.8 172.75 -0.05
300 299.95 -0.05
CLAMP CIIECK : GOOD
SMOOTHLY OF SLIDER CHECK : GOOD
THE DEVIATION OF ZERO POINT :
GAP BETWEEN MEASURING F'ACES
: OUTSIDE JAW NO CLEARANCE BETWEEN TIIE FACES.
: INSIDE JAW TTIE LIGIIT IS DISCERNIBLE BETWEEN THE FACES.
PARALLELISM OF EXTERNAL MEASURING FACES <0.05 film
THE T]NCERTAINTY OF MEASUREMENT WAS 0.065 ilrm
THE REPORTED UNCERTAINTY OF MEASI.]REMENT WAS BASED ON STANDARD
UNCERTAINTY MULTIPLIED BY A COVERAGE FACTOR K=2, PROVIDING A LEYEL OF
THE UNCERTAINTY OF APPROKMATELY 95o/o