CALIBRATION REPORT CERT NO. : DC19/0832
PAGE: 3 OF 3
R.F'SUI T O['CAI.IBRATTON : ADruSTMENT ( NO )
FTJNCTION: OUTSIDE MEASI.JREMENT
O
SCALE RANGE : mm TO 3OO
RESOLUTION: film
STANDARD LENGTH truC READING ERROR
( ttu" ) (mm) ( rntn )
ZERO 0.00 0.00
0.5 0.45 -0.05
1.4 1.35 -0.05
5.1 5.05 -0.05
7.7 7.65 -0.05
10.3 10.25 -0.05
22.9 22.85 -0.05
37.6 37.55 -0.05
70.2 70.15 -0.05
t22.8 722.75 -0.05
200 199.95 -0.05
300 299.95 -0.05
FIJNCTION : INSIDE MEASUREMENT
STANDARD LENGTH truC READING ERROR
(tnm) (rn-) (rnrn)
I D.b 15.50 0.00
16.4 16.40 0.oo
20.1 20.10 0.00
22.7 22.70 0.00
30.3 30.30 0.00
37.9 37.90 0.00
67.6 67.60 0.oo
95.2 95.20 0.00
t72.8 172.80 0.00
300 300.00 0.00
CLAMP CIIECK : GOOD
SMOOTHLY OF SLIDER CHECK: GOOD
THE DEYIATION OF ZERO POINT : 0.00
GAP BETWEEN MEASURING FACES
: OUTSIDE JAW NO CLEARANCE BETWEEN THE FACES.
: INSIDE JAW TTIE LIGHT IS DISCERNIBLE BETWEEN THE FACES.
PARALLELISM OF EXTERNAL MEASURING FACES 0.05 rnm
THE I.]I\ICERTAINTY OF MEASUREMENT WAS 0.065 mm
TT{E REPORTED UNCERTAINTY OF MEASUREMENT WAS BASED ON STANDARD
UNCERTAINTY MTILTIPLIED BY A COVERAGE FACTOR K=2, PROVIDING A LEVEL OF
THE UNCERTAINTY OF APPRO)flMATELY 95%
WC = I.JNIT UNDER CALIBRATE