CALIBRATION REPORT CERT NO. : DC19/0833
PAGE: 3 OF 3
RESUT T OF CAITBRATION : ADruSTMENT ( NO )
FT]NCTION: OUTSIDE MEASUREMENT
SCALE RANGE : 0 mm TO 3OO
RESOLUTION: O.O2 rlm
STANDARD LENGTH IruC READING ERROR
(oo") (-.n) (--)
ZERO 0.00 0.00
0.5 0.50 0.00
1.4 1.40 0.00
5.1 5.10 o.oo
7.7 7.70 0.00
10.3 10.30 0.00
22.9 22.90 0.00
37.6 37.60 0.00
70.2 70.20 o.oo
L22.8 t22.80 0.00
200 200.00 0.00
300 300.00 o.oo
FI.]NCTION : INSIDE MEASUREMENT
STAI{DARD LENGTH UUC READING BRROR
(nun) (nnn) (rrun)
15.5 15.50 0.00
16.4 16.40 0.00
20.1 20.10 0.00
22.7 22.70 0.00
30.3 30.30 0.00
37.9 37.90 0.00
67.6 67.60 0.00
95.2 95.20 0.00
172.8 172.80 0.00
300 300.00 0.00
CLAMP CIIECK: GOOD
SMOOTHLY OF SLIDER CHECK : GOOD
THE DEVIATION OF ZERO POINT : 0.00
GAP BETWEBN MEASURING T'ACES
: OUTSIDE JAW NO CLEARANCE BETWEEN THE FACES.
: INSIDE JAW THE LIGTIT IS DISCERNIBLE BETWEEN THE FACES.
PARALLELISM OF EXTERNAL MEASURING FACES <0.02 illm
TIIE T]NCERTAINTY OF MEASUREMENT WAS O.O27 ilrm
T}IE REPORTED UNCERTAINTY OF MEASI.]REMENT WAS BASED ON STANDARD
UNCERTAINTY MULTIPLIED BY A COVERAGE FACTOR K=2, PROVIDING A LEVEL OF
T}IE I.JNCERTAINTY OF APPRO)ilMATELY 9596
UUC = LINIT I]NDER CALIBRATE