Step 3 – connecting the applicators:
Warning: Applicators should be connected only when the system is turned OFF and the Power supply is not connected to an electrical outlet.
Have the back of the system facing you to unscrew the middle screw (identified in Fig. 11A) in the left-side back cover, using the supplied screwdriver.
Once unscrewed, simultaneously press the top and bottom clips of the left side back cover to be able to remove it altogether.
The uncovered panel reveals female type connectors to which the applicators would need to be plugged into (identified in Fig. 11B).
Note: the top-most female connector is disabled in this model, you may use any of the other three connectors.
Identify the male type connector at the end of the electrical cord of each applicator. Hold the connector with the bump on its circumference (Fig. 12C) directed at the matching dents in the female type connector and plug it inside all the way through until a “click” sound is heard. The insertion should be effortless if directed correctly (see Fig. 12A-B for the correct insertion direction and depiction of a well-connected applicator).
Fixate the electrical cords of the applicator in the grooves of the back panel to keep them sorted and organized as seen in Fig. 12D.
Fig. 12A-B Fig. 12C
Fig. 11A
Fig. 11B
Re-attach the left-side back cover and screw it back to place as in Fig. 11A above. See Fig. 12F for depiction of the GENEO X with a connected Oxygeneo applicator.