Function Checks
Chapter 2
Function Checks
1. Initial Actions have been performed.
Yes No
Go to “Initial Actions” on page 2-2.
2. Attach a secondary piece of equipment to the DISS gas outlet. The gas outlet
does not leak and is properly sealed.
Yes No
Go to RAP 2.1.
3. Attach a secondary piece of equipment to the Diamond®, Latch Index, or 2
Geometric quick connect key style gas outlet. The gas outlet does not leak and is
properly sealed.
Yes No
Go to RAP 2.2.
4. The gas outlet does not leak and is properly sealed when no secondary piece of
equipment is attached.
Yes No
For DISS gas outlets, go to RAP 2.1. For Diamond®, Latch
Index, or Geometric quick connect key style gas outlets go to
RAP 2.2.
5. Attach a secondary piece of equipment to the DISS, Diamond®, Latch Index, or
Geometric key style gas outlet. The secondary piece of equipment locks into
place and is properly attached to the gas outlet.
Yes No
Go to RAP 2.3.
6. For pressurized medical gases, remove the finish assembly from the rough-in
assembly. The rough-in assembly does not leak when the finish assembly is
removed.
Yes No
Go to RAP 2.4.
7. For vacuum outlets, ensure that the finish assembly is properly installed. The
rough-in assembly does not leak, and is properly sealed when the finish
assembly is installed.
Yes No
Go to RAP 2.5.
8. Go to “Final Actions” on page 2–4.
DiamondCare® Recessed Wall Outlet and Electrical Accessories (MAN01-045) Page 2-3