2.4 Rough-In Assembly Leakage (Pressurized Medical Gases)
1. Remove the finish assembly from the rough-in assembly on the gas outlet. The
rough-in assembly seals properly and does not leak.
Yes No
Replace the rubber bushing for all gas outlet types (refer to
procedure 4.3). If this solves the problem, go to “Final
Actions” on page 2-4. Otherwise, call BeaconMedæs
Technical Support at 1-(888) 4-MEDGAS (463-3427).
2. Go to “Final Actions” on page 2-4.
2
2.5 Rough-In Assembly Leakage (Vacuum Outlets Only)
1. Ensure the finish assembly is properly installed. The rough-in assembly does not
leak, and is properly sealed when the finish assembly is installed.
Yes No
Replace the rubber bushing in the rough-in assembly (refer to
procedure 4.4). If this solves the problem, go to “Final
Actions” on page 2-5. Otherwise, call BeaconMedæs
Technical Support at 1-(888) 4-MEDGAS (463-3427).
2. Go to “Final Actions” on page 2-4.