Medical Lubricated Vacuum


           4A.0  General Operation - TotalAlert Embedded Controls



           4A.4  Backup Vacuum Switch Set                      Adjusting Instructions
           Point Adjustments
                                                               1.  To make an adjustment, loosen the tamper
                                                                   resistant set screws (2) on the adjustment
           The backup switch is set at the factory to the
           operating point(s) as stated on the wiring diagram      sleeve.
           supplied with the unit. It is good practice to cycle   2.  Secure the hex body with an open-end
           the switch to determine actual operating points         wrench.  Hand turn the adjustment sleeve:
           before proceeding with readjustment.  Refer to          counter-clockwise to increase and clockwise to
           Figure 4A.4 for location of adjustment.
                                                                   decrease the set point.  The backup vacuum
                                                                   switch should always be set with falling
           CAUTION:
                                                                   vacuum level starting at a vacuum level higher
           •  ALWAYS change  vacuum setting gradually.             than the setpoint.

           •  ALWAYS check switch setting before making        3.  Using the vacuum gauge determine the
              any adjustments.                                     actuation point of the switch.
           •  DO NOT force adjustment sleeve when it
              becomes diffi  cult to turn.                     4. If the actuation point is above the desired
                                                                   value, turn the adjustment sleeve clockwise to
           •  ALWAYS isolate the vacuum transducer before          decrease the actuation point, and if it is below,
              making any adjustments to the backup vacuum          turn the adjustment sleeve counter-clockwise
              switch.                                              to increase it.
                                          Tamper Resistant
                         Increase         Set Screws (2)       5.  For exact vacuum setting, cycle vacuum switch
                                                                   and make fi ne adjustments by repeating steps
                                                                   2 through 4 (trial and error process) until the
                                                                   desired setting is obtained.

                                                               6.  Secure the tamper resistant set screws (2) on
                                                                   the adjustment sleeve.

                                                               CAUTION: Do not overtighten set screws.





                                         Decrease




           Figure 4A.4 Backup Vacuum Switch