Medical Lubricated Vacuum


           6.0  Maintenance


           6.2.3   Automotive-Type Oil Filter                  For 1.5 Hp, 2 Hp, 3 Hp and 5 Hp
                                                               5 Hp pump model RC101.  Incorporated into
           The pump is equipped with an automotive-type        horizontal tankmount system designs.
           oil fi lter.  (See Figure 6.1 for location.)
                                                               NOTE: Refer to Figure 6.2 for numbers referenced
           Replace the oil fi lter when changing the oil.  After   below in parenthesis.
           draining the oil, remove the oil fi lter and replace
           it with a new one using a BeaconMedæs oil fi lter.    To replace the exhaust fi lters, unscrew the screws
           Make sure to tighten the fi lter securely against the   (146) and lockwashers (143) from the exhaust
           aluminum sealing surface so that leaks will not     cover plate (145).  Remove cover plate and gasket
           occur.                                              (141).

           6.3   Drip Leg and Filters                          Use a slotted head screw driver to loosen the
                                                               exhaust fi lter retaining spring (125), then rotate
           6.3.1   Exhaust Drip Leg Valve                      and remove the spring.  Pull the fi lter  cartridge
                                                               (120) out of the exhaust box.
           Each pump should have a drip leg at the exhaust
           port on the pump.  This valve should be checked     Reinstall the fi lter elements.  Make sure the open
           daily at fi rst, then depending on the moisture      end of the element is properly seated down in its
           accumulated, could be checked less frequently.      recess in the exhaust box with the O-ring (121)
                                                               correctly positioned.  The indicating arrow on each
           6.3.2   Inlet Filter                                fi lter element must be pointing upward toward
                                                               the top of the exhaust box after the assembly has
           Each pump is equipped with an inlet fi lter.  The    been installed.
           fi lter cartridge should be checked on a monthly
           basis and cleaned as required.  The fi lter cartridge   Retain the fi lter with the spring clip (125), tighten
           should be changed annually.                         the tension screw until the fi lter is secure.

           To clean the inlet fi lter, unsnap the lid clamps    Place the exhaust port gasket (141) and cover
           and lift off  the fi lter lid.  Remove cartridge, being   (145) in position on the exhaust box and retain
           careful not to knock any foreign particles present   with the cap screws (146) and lockwashers (143).
           inside the canister into the pump section.  Clean
           foreign particles from the canister with an air hose,   For 5 Hp, 7.5 Hp, and 10 Hp
           and carefully back fl ush the fi lter cartridge with   5 Hp pump model RC0155.  Incorporated into
           shop air.  If the fi lter cartridge has been subjected   basemount system designs.
           to moisture or is extremely dirty, it may need
           replacement.                                        NOTE: Refer to Figure 6.3 for numbers referenced
                                                               below in parenthesis.
           6.3.3   Exhaust Filter
                                                               To replace the exhaust fi lters, unscrew the screws
           Exhaust fi lters should be checked monthly.  A       (142) and washers (143) from the exhaust cover
           pressure gauge is supplied as part of the oil fi ll   plate (140).  Remove cover plate and gasket (141).
           plug.  This gauge has a green fi eld and a red fi eld.
           A pressure within the green fi eld indicates normal   Remove the distance pin (137), the baffl  e strainer
           pressure.  Any pressure in the red fi eld (for a     (130) and the sheet metal plate (136).
           continuous period of time) requires an immediate
           change of the exhaust fi lter(s).  Annually, or as   Remove the discharge fi lter assembly by unscrewing
           necessary, replace the exhaust fi lter elements.     the nut (134) and removing the lockwasher (132).