Medical Lubricated Vacuum


           6.0  Maintenance


           6.5  General Inspections


           6.5.1   Monthly Inspections

           A general inspection should be performed on a
           regular basis (monthly) for safety items.  Items to
           inspect include all wiring, fl ex hoses, and other
           items.  If a damaged item is viewed, call your local
           BeaconMedæs service technician for a thorough
           inspection and report of fi ndings.

           6.5.2   Every Six Months


           A thorough inspection  of the vacuum pump
           cooling air intake and discharge grating should
           be performed at least every six months or more
           frequently if conditions require.  If a dust/dirt
           buildup is visible, clean the grating to remove the
           buildup.  Excess dust/dirt buildup in these areas
           will prevent air from cooling the vacuum pump,
           aff ecting peformance of the vacuum system.

           6.5.3   Oil Leak


           If upon inspection of the vacuum system, a
           noticeable amount of oil is seen on the base of
           the system or on the fl oor around the system,
           immediate action is required to prevent the
           possibility of injury.  Using an oil absorbent or like
           product, absorb the oil on the base frame or fl oor
           to remove the potential of slipping.  Call your local
           BeaconMedæs service technician for a thorough
           inspection of the vacuum pumps.