Page 36 - Med Plaza and Cancer Center
P. 36

ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

           Facility   Florida Hospital Orlando-Med Plaza      Inspector  Steve Gourley
          City/State  Orlando, FL                             Date       12-4-17         Page       1 of 2
          Contractor  n/a                                     Facility Contact   Darrell Huckaby

          Area Served   OR 4                                  Floor/Bldg   1  Floor Med Plaza
                                                                           st

           Zone Valve Location for this Area   Outside OR4            ZVB Manufacturer        Hill-Rom

         FL Hosp. Tag #          No FL Hosp #
          Location IAW NFPA 99   Pass               Labeled      Pass     Gauges     Pass      Leak Test   Pass
           Test Gauge   O2    52    AIR     49   VAC    22   N2O    53    N2    175  WAGD     22
           Pressure vs.
           ZVB Gauge          51           50           22          51          179           22

           Valve Size      3/4          3/4          1           3/4         1/2          3/4
          Comments:

         Gas/Pipeline Purity  Oxygen       Med Air        N2O            Comments
         Hydrocarbons               <1 PPM        <1 PPM          <1 PPM
         Dew Point                     -46             23            -27
         Carbon Monoxide            <1 PPM        <1 PPM          <1PPM
         Carbon Dioxide             <1 PPM       370 PPM          <1 PPM
         Gas Concentration         99.9%O2     21.0 (%O2)      99.9%N2O

         Particulates                   <1             <1             <1
         Odor/Discoloration           None          None            None
         Test Location                OR4            OR4            OR4

         Area Alarm Panel Location for this Area         Across from decontamination room
         Location IAW NFPA 99       pass                 Labeled    Pass   Gauges     Pass     Leak Test   Pass
         Area Alarm Panel Manufacture Data         Hill rom Med Plus (GP-L-5PS-01-6G-001)
                     GAUGE     LOW      HIGH                        TEST
          GAS/VAC      PSI    ALARM    ALARM    VISUAL   AUDIBLE   SWITCH   SENSOR        COMMENTS        PASS/FAIL
            OXY        51       40       60       P        P         P         SS                           PASS
             AIR       50       40       60       P        P         P         SS                           PASS
            VAC        25       12        -       P        P         P         SS                           PASS
            N2O        53       40       60       P        P         P         SS                           PASS
             N2        170      130      200      P        P         P         SS                           PASS
            WAGD       24       12        -       P        P         P         SS                           PASS
         PS= Patient Side   SS= Source Side   NW= Not Working   CNT= Could Not Test   NA = Not Applicable    NP = Not Present