ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

           Facility   Florida Hospital Orlando-Med Plaza      Inspector   Steve Gourley
          City/State  Orlando, FL                               Date        11-30-17     Page       1 of 2
          Contractor  n/a                                       Facility Contact  Darrell Huckaby

          Area Served   Recovery Beds 6-9                       Floor/Bldg   1  Floor Med Plaza
                                                                              st

           Zone Valve Location for this Area   Outside recovery beds 6-9   ZVB Manufacturer   Beacon

         FL Hosp. Tag #          (O2/Air) #000000227014, (N2O/VAC) # 000000227016
          Location IAW NFPA 99   Pass               Labeled      Pass        Gauges   Pass     Leak Test   Pass
           Test Gauge   O2    52    AIR    49   VAC     28   N2O     N/A
           Pressure vs.
           ZVB Gauge          53           51           28           N/A

           Valve Size      3/4          1/2         11/2          3/4
          Comments:


         Gas/Pipeline Purity   Oxygen      Med Air        Comments
         Hydrocarbons               <1 PPM        <1 PPM
         Dew Point                     -46             23
         Carbon Monoxide            <1 PPM        <1 PPM
         Carbon Dioxide             <1 PPM       370 PPM
         Gas Concentration         99.9%O2     21.0 (%O2)
         Particulates                   >1             >1
         Odor/Discoloration     None/Light      None/Light
         Test Location               Bed 7          Bed 7


         Area Alarm Panel Location for this Area         In nurse’s station for beds 6-9
         Location IAW NFPA 99       Pass                 Labeled    Fail   Gauges     Pass     Leak Test   Pass
         Area Alarm Panel Manufacture Data         Ohio black face (GP-L-5PS-01-4G-002)
                     GAUGE     LOW      HIGH                        TEST
          GAS/VAC      PSI    ALARM    ALARM    VISUAL   AUDIBLE   SWITCH   SENSOR        COMMENTS        PASS/FAIL
            OXY        53       40       60       P        P         P         PS      Normal light is out.   Pass
             AIR       51       40       60       P        P         P         PS                           Pass
            VAC        25       12        -       P        P         P         PS                           Pass
           PS= Patient Side   SS= Source Side   NW= Not Working   CNT= Could Not Test   NA = Not Applicable    NP = Not Present



              Corporate Office ∙ 4518 Clements Rd. Lakeland Florida 33811 Phone: 863-646-7555 ∙ Fax 863-646-2999
                                 “Medical Gas Pipeline Equipment Service, Sales & Verifications”