ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

           Facility   Florida Hospital Orlando-Med Plaza      Inspector  Steve Gourley
          City/State  Orlando, FL                             Date       12-4-17         Page       1 of 2
          Contractor  n/a                                     Facility Contact   Darrell Huckaby

                                                                           th
          Area Served   OR 4                                  Floor/Bldg   4  Floor Med Plaza

                                              Outside OR4
           Zone Valve Location for this Area   door frame# K-5PN-04-E-85-&   ZVB Manufacturer           Medaes
         FL Hosp. Tag #          O2/MA/VA/NO/N2/CO-P-5PN-04-B-Z-W-04
          Location IAW NFPA 99   Pass               Labeled      Pass     Gauges     Pass      Leak Test   Fail
           Test Gauge   O2    52    AIR    49   VAC     24   N2O    52    N2    169   CO2     51
           Pressure vs.
           ZVB Gauge          52           49           25          51          175           51

           Valve Size
          Comments: Gauges leak, N2 reading high

         Gas/Pipeline Purity  Oxygen       Med Air        N2O            CO2             Comments
         Hydrocarbons               <1 PPM        <1 PPM          <1 PPM         <1 PPM
         Dew Point                     -46             23            -27             -17
         Carbon Monoxide            <1 PPM        <1 PPM          <1PPM          <1 PPM
         Carbon Dioxide             <1 PPM       370 PPM          <1 PPM       5000PPM
         Gas Concentration         99.9%O2     21.0 (%O2)      99.9%N2O       99.9%CO2

         Particulates                   <1             <1             <1             <1
         Odor/Discoloration           None          None            None           None
         Test Location                OR4            OR4            OR4             OR4

         Area Alarm Panel Location for this Area         Behind OR nurses station
         Location IAW NFPA 99       Pass                 Labeled    Pass   Gauges     Pass     Leak Test   Pass
         Area Alarm Panel Manufacture Data         Medaes Analog (GP-L-5PC-04-5G-SA4) VAC Alarm (GP-L-5PC-04-6-G-Z03)
                     GAUGE     LOW      HIGH                        TEST
          GAS/VAC      PSI    ALARM    ALARM    VISUAL   AUDIBLE   SWITCH   SENSOR        COMMENTS        PASS/FAIL
            OXY        52       40       60       P        P         P         SS                           PASS
             AIR       51       40       60       P        P         P         SS                           PASS
            VAC        23       12        -       P        P         P         SS                           PASS
            N2O        54       40       60       P        P         P         SS                           PASS
             N2        173      130      200      P        P         P         SS                           PASS
            CO2        52       40       60       P        P         P         SS                           PASS
           PS= Patient Side   SS= Source Side   NW= Not Working   CNT= Could Not Test   NA = Not Applicable    NP = Not Present