ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

           Facility   Florida Hospital Orlando-Med Plaza      Inspector  Anthony Voight
          City/State  Orlando, FL                             Date       12-4-17         Page       1 of 2
          Contractor  n/a                                     Facility Contact   Darrell Huckaby

          Area Served   OR 7                                  Floor/Bldg   4  Floor Med Plaza
                                                                           th

           Zone Valve Location for this Area   Outside OR7 (door frame# K-5PC-04-f-70-8)   ZVB Manufacturer   Medaes
         FL Hosp. Tag #          O2/MA/VA/NO/N2/CO-P-5PC-04-B-Z-W-07
          Location IAW NFPA 99   Pass               Labeled      Pass     Gauges     Pass      Leak Test   Pass
           Test Gauge   O2    54    AIR    52   VAC    -25   N2O    53    N2    170   CO2     53
           Pressure vs.
           ZVB Gauge          52           49          -25          54          180           54

           Valve Size       ½           ½            1           ½            ½           1/2
          Comments: N2 missing Label


         Gas/Pipeline Purity  Oxygen       Med Air        N2O            CO2             Comments
         Hydrocarbons               <1 PPM        <1 PPM          <1 PPM         <1 PPM
         Dew Point                     -46             23            -27             -17
         Carbon Monoxide            <1 PPM        <1 PPM          <1PPM          <1 PPM
         Carbon Dioxide             <1 PPM       370 PPM          <1 PPM       5000PPM
         Gas Concentration         99.9%O2     21.0 (%O2)      99.9%N2O       99.9%CO2
         Particulates                   <1             <1             <1             <1
         Odor/Discoloration           None          None            None           None

         Test Location                OR7            OR7            OR7             OR7

         Area Alarm Panel Location for this Area         In OR nurses station

         Location IAW NFPA 99       Pass                 Labeled    Pass   Gauges     Pass     Leak Test   Pass
         Area Alarm Panel Manufacture Data         Medaes Analog (GP-L-5PC-04-5GSA4)
                     GAUGE     LOW      HIGH                        TEST
          GAS/VAC      PSI    ALARM    ALARM    VISUAL   AUDIBLE   SWITCH   SENSOR        COMMENTS        PASS/FAIL
            OXY        54       40       60       P        P         P         SS                           PASS
             AIR       51       40       60       P        P         P         SS                           PASS
            VAC        27       12        -       P        P         P         SS                           PASS
            N2O        51       40       60       P        P         P         SS                           PASS
             N2        170      130      200      P        P         P         SS                           PASS
            CO2        54       40       60       P        P         P         SS                           PASS
           PS= Patient Side   SS= Source Side   NW= Not Working   CNT= Could Not Test   NA = Not Applicable    NP = Not Present

              Corporate Office ∙ 4518 Clements Rd. Lakeland Florida 33811 Phone: 863-646-7555 ∙ Fax 863-646-2999
                                 “Medical Gas Pipeline Equipment Service, Sales & Verifications”
                                                      ©Daniel Voight 2009