                   Preventive Maintenance
                   Chapter 6: General Procedures


                   Preventive Maintenance




                                 WARNING:
                                 Only facility-authorized personnel should perform preventive
                                 maintenance on the MedPlus TotalAlert™ Alarm Network. Preventive
                                 maintenance performed by unauthorized personnel could result in
                                 personal injury or equipment damage.

                                 The MedPlus TotalAlert™ Alarm Network requires an effective maintenance
                                 program. We recommend that you perform bi-annual preventive maintenance
                                 (PM) and testing for Joint Commission on Accreditation of Healthcare
                                 Organizations (JCAHO). PM and testing not only meet JCAHO requirements
                                 but will help ensure a long, operative life for the MedPlus TotalAlert™ Alarm
                                 Network. PM will minimize downtime due to excessive wear.

                                 The following PM schedule guides you through a normal PM procedure on the
                                 MedPlus TotalAlert™ Alarm Network. During this PM process, check each
                                 item on the schedule, and make the necessary adjustments.

                                 Two effective ways to reduce downtime are keeping close records and
                                 maintaining the MedPlus TotalAlert™ Alarm Network.


                                 The MedPlus TotalAlert™ Alarm Network is designed to operate reliably with
                                 a minimum amount of scheduled maintenance. Components, such as the LEDs,
                                 and the vacuum fluorescent display, are used to provide visual indication and
                                 have been chosen for their extended service life. As a result, no replacement of
                                 components is necessary unless a problem is identified during regularly
                                 scheduled testing.


                                 When the MedPlus TotalAlert™ Alarm Network Conversion Kit—Area
                                 Alarms have transducers inside the rough-in box, check the hoses on the
                                 transducer assemblies annually for cracking, wear, or leakage.
























                   Page 6 - 30           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)