PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                                                                                    Preventive Maintenance
                                                                              Chapter 6: General Procedures

                       Preventive Maintenance Schedule


                              Permanent records of all test results for alarm systems should be made at the
                              time of installation and at all required intervals as specified by NFPA 99 (see
                              table 6-1 on page 6-31).


                                            Table 6-1.  Preventive Maintenance Schedule

                                Function                                     Procedure
                 Gas pressures                              Check the readings on the area alarm module
                                                            digital displays daily.
                 Audible alarm signals                      Test once a month by pressing the test keys on
                                                            each front panel (see “Master Alarm Self-Test”
                                                            on page 2-5), (see “Area Alarm Self-Test” on
                                                            page 2-6), and (see “Local Alarm Self-Test” on
                                                            page 2-7).
                 Visual alarm signals                       Test once a month by pressing the test keys on
                                                            each front panel (see “Master Alarm Self-Test”
                                                            on page 2-5), (see “Area Alarm Self-Test” on
                                                            page 2-6), and (see “Local Alarm Self-Test” on
                                                            page 2-7).
                 Transducer visual indicators               Inspect once a month (see “Transducer” on
                                                            page 2-11).
                 Alarm system components                    Check every alarm component for wear.
                 Power and signal wiring                    Inspect all wiring for good condition.
                 Electrical connections                     Check all electrical connections for tightness,
                                                            and tighten as needed.
                 Gas hoses  (on local-mount transducers)    Inspect hoses for cracking, wear, collapse, or
                                                            leakage annually.
                 General appearance                         Check cleanliness. Clean if necessary, see
                                                            “Cleaning” on page 6-28.



                     Contact Hill-Rom Technical Support at (800) 445-3720 for information on having these tests
                                                                                                                 6
                     performed by a Hill-Rom Service Representative.