Page 346 - Womens Pavilion
P. 346

Oil-Less Claw Medical Vacuum


            Appendix B: Variable Speed Drive Inverter


            B.5  Cooling Fan Replacement (continued)


            Reinstallation

            1. After confi rming the orientation of the fan, reinstall the fan so that the arrow on the left of “AIR
               FLOW” faces up.





                           AIR FLOW


                   <Fan side face>

            CAUTION: Installing the fan in the opposite air fl ow direction can cause the inverter life to be shorter.

            2.  Reconnect the fan connectors.  When wiring, use care to avoid the cables being caught by the fan.
















               FR-F720-00250                 FR-F720-00340 to 00490
               FR-F740-00126                 FR-F740-00170 to 00250


            3.  Reinstall the fan cover.

            2.  Insert hooks until                      1. Insert hooks
               you hear a click  1.  Insert hooks into      into holes.    2.  Insert hooks until
               sound.              holes.                                      you hear a click
                                                                               sound.










               FR-F720-00250                            FR-F720-00340 to 00490
               FR-F740-00126                            FR-F740-00170 to 00250






                                                             B-27                                 4107 9000 95.04
   341   342   343   344   345   346   347   348   349   350   351