Page 347 - Womens Pavilion
P. 347
Oil-Less Claw Medical Vacuum
Appendix B: Variable Speed Drive Inverter
B.6 Inverter Replacement
WARNING:
Before starting inverter replacement,
switch power off , wait for at least 10
minutes, and then check the voltage
with a tester and such to ensure
safety.
The inverter can be replaced with the control circuit wiring kept connected. Before replacement, remove
the wiring cover of the inverter.
1. Loosen the two installation screws in both ends of the control circuit terminal block. (These screws
cannot be removed.) Pull down the terminal block from behind the control circuit terminals.
2. Using care not to bend the pins of the inverter’s control circuit connector, reinstall the control circuit
terminal block and fi x it with the mounting screws.
4107 9000 95.04 B-28