Troubleshooting





                   SYMPTOM                                CAUSE                        CORRECTIVE ACTION

          3.  Power light is ON, however     a.  Loose ribbon cable from LCD   i.  Ensure that the cable is firmly in
             there is no display on LCD         screen to board                    it’s socket on the LCD screen and
             screen                                                                board


                                                                               ii.  Replace the LCD module

          4. No audible alarm                a. DC power cable is disconnected,   i.  Ensure that the DC power cable
                                                loose or check ribbon cable        from the system power supply
                                                                                   is  connected to the LCD module
                                                                                   snugly


                                                                               ii. Replace  LCD board



          5. Audible signal will not silence  a. Faulty display module         i.  Disconnect the ribbon cable from
                                                                                   the back of the faulty display module
                                                                                   and replace the LCD module


                                             b. Connection of the DC power     ii. Disconnect  the  DC  power cable
                                                cable from system power            from the LCD module and then
                                                supply to LCD module is loose      reconnect.  If  audible alarm  still
                                                                                   persists, replace the System Power
                                                                                   Supply unit


                                             c. Faulty push button             iii. Replace the LCD module