Troubleshooting





                   SYMPTOM                                CAUSE                        CORRECTIVE ACTION
          6.  Gas reading incorrect         a.  Loose connection of DISS       i.  Ensure that the sensor module is
                                                fittings                           properly connected to the DISS
                                                                                   demand check-valve


                                            b.  Sensor module is not properly   ii.  Ensure that the sensor module is
                                                wired to the display module        properly wired to the LCD module
                                                                                   by using wiring diagram in Appendix
                                                                                   D or Appendix E



                                            c.  Defective sensor or requires   i.  Replace the sensor module
                                                calibration


                                            d.  Defective LCD display module   ii.  Replace the LCD display module



          7.  Display shows “NO SENSORS”    a.  No sensor(s) are connected     i.  Make  sure sensor module(s)  are
                                                to the LCD board                   connected to LCD board (see
                                                                                   Appendix  D and Appendix E).

                                            b.  Program not set up             ii.  Press set-up & select button to
                                                                                   program all connected sensors (see
                                                                                   LCD Display Set-Up on pg. 10)

                                            c.  Faulty sensors                 iii.  Replace sensors






            FACTORY DEFAULT SETTING GAS





                   Mid Pressure         Hi      =  60 psi
                                        Low     =  40 psi




                   Vacuum               Hi      =  32 inHg
                                        Low     =  12 inHg




                   High Pressure        Hi      =  195 psi
                                        Low     =  140 psi