Troubleshooting
SYMPTOM CAUSE CORRECTIVE ACTION
6. Gas reading incorrect a. Loose connection of DISS i. Ensure that the sensor module is
fittings properly connected to the DISS
demand check-valve
b. Sensor module is not properly ii. Ensure that the sensor module is
wired to the display module properly wired to the LCD module
by using wiring diagram in Appendix
D or Appendix E
c. Defective sensor or requires i. Replace the sensor module
calibration
d. Defective LCD display module ii. Replace the LCD display module
7. Display shows “NO SENSORS” a. No sensor(s) are connected i. Make sure sensor module(s) are
to the LCD board connected to LCD board (see
Appendix D and Appendix E).
b. Program not set up ii. Press set-up & select button to
program all connected sensors (see
LCD Display Set-Up on pg. 10)
c. Faulty sensors iii. Replace sensors
FACTORY DEFAULT SETTING GAS
Mid Pressure Hi = 60 psi
Low = 40 psi
Vacuum Hi = 32 inHg
Low = 12 inHg
High Pressure Hi = 195 psi
Low = 140 psi