Rough-In Assembly Leakage
                                                                                                 Chapter 2

                2.4    Rough-In Assembly Leakage (Pressurized Medical Gases)

                       1. Remove the finish assembly from the rough-in assembly on the gas outlet. The
                          rough-in assembly seals properly and does not leak.
                                  Yes   No
                                        Replace the rubber bushing for all gas outlet types (refer to

                                            procedure 4.3). If this solves the problem, go to “Final
                                            Actions” on page 2-4. Otherwise, call BeaconMedæs
                                            Technical Support at 1-(888) 4-MEDGAS (463-3427).


                       2. Go to “Final Actions” on page 2-4.

                                                                                                                2
                2.5    Rough-In Assembly Leakage (Vacuum Outlets Only)

                       1. Ensure the finish assembly is properly installed. The rough-in assembly does not
                          leak, and is properly sealed when the finish assembly is installed.
                                  Yes   No
                                        Replace the rubber bushing in the rough-in assembly (refer to

                                            procedure 4.4). If this solves the problem, go to “Final
                                            Actions” on page 2-5. Otherwise, call BeaconMedæs
                                            Technical Support at 1-(888) 4-MEDGAS (463-3427).


                       2. Go to “Final Actions” on page 2-4.