Page 20 - 3-PLY Level 2 Hensler Mask Booklet - ALL
P. 20
REPORT NO.: HW20200322024S
EN 14683: 2019
Clause Requirement -Test Result• Remark Verdict
When tested in accordance with Annex B, the p
bacterial filtration efficiency ( BFE ) of the medical face
mask shall conform to the minimum value given for
the relevant tvo e in Table 1.
Test Type la Type II Type IIR
Bacterial
filtration .: 95 .: 95 .: 95
efficiency
( BFE ) , (%)
Differential
pressure < 49, 0 < 49, 0 < 49, 0
( Pa/cm2 )
Splash
5.2.2
resistance Not Not .: 16,0
pressure required required
l kPa l
Microbial
cleanlines s 30 s 30 s 30
s ( cfu/ q)
a Type I medical face masks should only be used
for patients and other persons to reduce the risk of
spread of infections particularly in epidemic or
pandemic situations. Type I masks are not
intended for use by healthcare professionals in an
operating room or in other medical settings with
similar requirements
When tested in accordance with Annex C, the p
differential pressure of the medical face mask shall
5.2.3
conform to the value given for the relevant type in
Table 1.
When tested in accordance with ISO 22609 the p
resistance of the medical face mask to penetration of
5.2.4 splashes of liquid shall conform to the minimum value
aiven for Tvoe IIR in Table 1.