REPORT NO.: HW20200322024S

 EN 14683: 2019
 Clause   Requirement -Test   Result• Remark  Verdict
 When tested in accordance with Annex B, the   p
 bacterial filtration efficiency  ( BFE )  of the medical face
 mask shall conform to the minimum value given for
 the relevant tvo e in Table 1.
 Test   Type la   Type II   Type IIR
 Bacterial
 filtration   .: 95   .: 95   .: 95
 efficiency
 ( BFE ) , (%)
 Differential
 pressure   < 49, 0   < 49, 0   < 49, 0
 ( Pa/cm2 )
 Splash
 5.2.2
 resistance   Not   Not   .: 16,0
 pressure   required   required
 l kPa l
 Microbial
 cleanlines   s 30   s 30   s 30
 s  ( cfu/ q)
 a Type I medical face  masks should only be used
 for patients and other persons to reduce the risk of
 spread  of  infections  particularly  in  epidemic  or
 pandemic  situations.  Type  I  masks  are  not
 intended for use by  healthcare professionals in an
 operating  room  or  in  other  medical  settings  with
 similar requirements
 When tested in accordance with Annex C, the   p
 differential pressure of the medical face mask shall
 5.2.3
 conform to the value given for the relevant type in
 Table 1.
 When tested in accordance with ISO 22609 the   p
 resistance of the medical face mask to penetration of
 5.2.4   splashes of liquid shall conform to the minimum value
 aiven for Tvoe IIR in Table 1.