Page 33 - SRL Diagnostics
P. 33

Do’s                                  Don’ts

                         If excessive mucus is present, this   DO NOT routinely clean the cervix
                         should be gently removed before       or take a cervical swab before
                         sampling                              taking the sample

                         Use either the Cervex Brush on its    DO NOT use an endocervical
                         own or a plastic spatula and          brush in isolation
                         endocervical brush combination.
                                                               DO NOT use a wooden spatula
                         The Cervex brush should be            under any circumstances
                         rotated 5 times in clockwise
                         direction. The plastic spatula        DO NOT use lubricant
                         should be rotated through 360
                         degrees and the endocervical          DO NOT leave the collection
                         brush rotated through one quarter     device in the vial whilst dealing
                         to one half turn.                     with the pt.

                         Immediately rinse the collected       DO NOT over tighten the lid on the
                         material into the vial.  Replace the   vial
                         lid and tighten so that black torque
                         line on the cap passes the black      DO NOT detach the head of the
                         torque line on the vial to avoid      Cervex brush in the vial
                         leakage.
                                                               DO NOT place multiple labels on
                         Keep the unlabelled portion of the    the vial
                         sample vial free so that the
                         contents can be seen.                 DO NOT apply bar coded labels
                                                               vertically on the vial
                         If bar coded labels are used these
                         must be applied horizontally          DO NOT use expired vials
                         around the vial
                                                               DO NOT delay sending the vials to
                         Samples should be sent to the lab     the lab
                         without delay