                   Another study tested the Overhand Throw, the Duncan loop, the Revo knot, the Tennessee Slider, and the
                   SMC  knot  (20  ).  All  knots  were  tied  using  braided,  nonabsorbable  suture.  Under  cyclic  loading,  the
                   Overhand throw displaced significantly greater than the other knots. The Revo knot was shown to have
                   the  highest  load  to  clinical  failure,  followed  by  the  Tennessee  Slider  and  the  SMC  knot,  which  were
                   statistically similar, the Duncan loop, and then the Overhand Throw, which had the lowest load to clinical
                   failure (Fig. 47 ). Concerning ultimate failure, the Revo knot and the SMC knot sustained the highest loads.
                   Again, the Overhand throw was able to tolerate the least load to failure.























                     FIGURE 47. Load to Clinical Failure (20).

                     Significance  of  the  internal  locking  mechanism  for  loop  security  enhancement  in  the  arthroscopic
                     knot. Arthroscopy, 17(8): 850-5, 2001.