Page 59 - PR 2014 2016 07 Nuclear Science and Technology
P. 59

Nuclear Science and Technology | Progress Report  201





               Scientific dosimeters for all phantoms. For the   given by IMRT planning and the repeatability
               three phantoms studied, the dose response      of TL responses is better than + 4.12%, lower
               curves to 12 MeV electrons presented a linear   than 5% acceptable for radiation therapy. The
               behavior for doses from LDL up to 5 Gy. All    scattered radiation doses received by structures
               repeatability values are better than +3.5%.    3, 4 and 5 corresponded on average to 16.14%
               CaSO :Dy dosimeters showed maximum vari-       of the highest dose received by the structure
                    4
               ation of TL sensitivity relative to  Co of 11.6%   1, according to the planning. The LiF:Mg,Ti do-
                                              60
               between solid water and PMMA phantoms and      simeters demonstrated to have good accuracy
               microLiF:Mg,Ti dosimeters showed maximum       in all measures of IMRT planning. Comparing
               variation of TL sensitivity relative to  Co of   the doses calculated by Al O :C  dosimeters
                                                    60
                                                                                         2  3
               9.5% to liquid water and solid water phantoms.   using OSL technique with the doses provided
               According to the results, the phantom mate-    by the planning system, it was observed that
               rial affected the electron spectrum incident   the dose for the cavity 1 was underestimated
               on the detector and altered the response of
               the dosimeters to 12 MeV clinical electron                                  Figure 35. PMMA
               beam using microLiF:Mg,Ti and CaSO :Dy as                                   phantom developed
                                                    4                                      to IMRT technique
               thermoluminescent detectors.                                                quality assurance


               TL and OSL dose response using a PMMA
               phantom for IMRT technique quality assurance


               The principle of IMRT is to treat a patient from
               a number of different directions (or continuous
               arcs) with beams of non-uniform fluences,      Three-dimensional dosimetric verification
               which have been optimized to deliver a high    system using Fricke gel solution applied to
               dose to the target volume and an acceptably    verification of Volumetric Modulated Arc
               low dose to the surrounding normal structures.   Radiation Therapy (VMAT) in the treatments
               This study had as objective to provide informa-  with respiration target movement
               tion to the physicist regarding the application
               of different dosimeters type, phantoms and     Volumetric Modulated Arc Therapy (VMAT) is
               analysis technique for Intensity Modulated     one of the methods most commonly used in
               Radiation Therapy (IMRT) dose distributions    teletherapy to treat cancer. The various tech-
               evaluation. The measures were performed        nological advances and the evolution of treat-
               using dosimeters of LiF:Mg,Ti and Al O :C eval-  ment techniques made the VMAT as one of the
                                                 2  3
               uated by techniques of thermoluminescent (TL)   state of the art methods for the treatment of
               and Optically Stimulated Luminescence (OSL).   some cancers. Such advances require quality
               A polymethylmethacrylate (PMMA) phantom        control of these tools in order to ensure that
               with five cavities, two principal target vol-  the entire treatment process is satisfactory
               umes considered like tumors to be treated and   and accurate. To date, the community lacks
               other three cavities to measure the scattered   an experimental system capable of evalu-
               radiation dose was developed to carry out the   ating, considering the uncertainty levels, if
               measures Fig. 35. The doses evaluated to the   the computerized planning systems are able
               tumor simulators cavities using LiF:Mg,Ti do-  to consider the movement of targets in the
               simeters corresponding to the estimated doses   treatments with VMAT. In this work, it was
   54   55   56   57   58   59   60   61   62   63   64