208   Nuclear Science and Technology | Progress Report























               Figure 46 – Alanine/EPR dosimetric system: (a) Bruker ESR spectrometer (X-band); (b) alanine en-
               capsulated in polyethylene tube and buildup cap in polymethylmethacrylate.


               in gamma rays, X-rays and electron beams       glasses, sugar, ceramic and pharmaceuticals.
               sources (Fig. 46a).
                                                              Radioprotection Level
               The amino acid alanine is tissue-equivalent
               material with several features for dosimetric   An improvement on the calibration proce-
               applications: low fading, low uncertainty      dure in high exposure rate field was imple-
               (<3%), no dose rate dependence and non-de-     mented for portable survey meters used in
               structive EPR signal readout.                  radioprotection routines. These equipments
                                                              were tested in relation to their response as:
               For therapeutic dose range, the High Dose      repeatability, reproducibility, and exposure
               Dosimetry Laboratory of IPEN developed a       rate measurements. According to the cali-
               dosimetric system based on alanine/EPR. The    bration results obtained in a Cs-137 gamma
               detector is encapsulated in special polyeth-   radiation field, the procedure for portable
               ylene tube that reduces the humidity prob-     survey meters used in a high exposure rates
               lems and improves the mechanical resistance.   was established at Calibration Laboratory
               A computer program to extract signals from     of IPEN-CNEN/SP. It was also introduced an
               noise spectra based on the wavelet transform   improved test program to be applied before
               was developed in order to allow the use of     the calibration procedure starts in a gamma
               the dosimetric system at radiotherapy dose     radiation field. Several problems related to
               ranges (1 to 20 Gy).                           the equipment operation as: check of the
                                                              battery voltage, display malfunction, could
               For industrial applications, the dosimetric    be avoided with these actions. About 15%
               system is based in commercial alanine pellets   of the equipments have shown this type
               which offer wide dose range from 0,01 to 100   of  problems  during  the  calibration  pro -
               kGy (Fig. 46b).                                cedures. The qualification test was made
                                                              with Geiger-Müller detectors, from differ-
               EPR spectrometry is powerful instrument        ent manufacturers and models, according
               for radiation dosimetry accident evaluation.    to energy dependence in standard field of
               Inorganic or organic substances present in     gamma radiation. The qualification results
               the accident scenario are candidates for ret-  obtained from two hundred detectors agree
               rospective dose study by the EPR technique.    with the national and international standard
               These items included nails, plastics, clothes,   recommendations. The data sheets and all