Page 67 - PR 2014 2016 07 Nuclear Science and Technology
P. 67

Nuclear Science and Technology | Progress Report  209





                                                                                           60
               information in relation to the new proce-      be calibrated, using gamma ( Co) or low en-
               dures were integrated in one database. The     ergy X radiation. As a reference system, a
               improvements implemented have reduced          secondary standard ionization chamber is
               costs and increased productivity.              used, traceable to the Physikalisch-Technische
                                                              Bundesanstalt, PTB, Germany, and to the Na-
               Diagnostic Radiology                           tional Laboratory of Metrology of Ionizing

                                                              Radiation, Brazil.
               It was developed the improvement of one
               calibration procedure of mammography do-       Instruments used in diagnostic radiology
               simeters (ionization chambers) to be applied   measurements can be tested in X radiation
               at clinical X-ray system of the calibration    qualities, using a Seifert X radiation system
               laboratory (IPEN-CNEN/SP). The clinical X-ray   (160 kV) and a reference system with four
               system was studied in their behavior, or test-  ionization chambers for diagnostic radiology
               ed with respect to their response such as:     measurements (mammography, computed
               repeatability, reproducibility of the nomi-    tomography, fluoroscopy, radiation protec-
               nal voltage, current and air kerma rate. The   tion and conventional diagnostic radiology
               results obtained show the possibility to use   ionization chamber) traceable to the PTB,
               the clinical X-ray system in calibration pro-  Germany (Fig. 47).
               cedures. After those studies, one calibration
               procedure of mammography dosimeters (ion-
               ization chambers) was established at calibra-
               tion laboratory.

                Metrology in Diagnostic Radiology



               The Calibration Laboratory has since 1980
               been calibrating instruments used in radia-
               tion protection and therapy measurements
               and belonging to hospitals, industries, clinics
               and other users located in São Paulo and in
               other parts of Brazil.


               Since 2000, calibration service is being offered
               to users of diagnostic radiology instruments
               with the establishment of standard radiation   Figure  47 -  X radiation calibration system.
               quality at this level.
                                                              The types of instruments calibrated are: sev-
               At the radiation protection level, there are   eral kinds of ionization chambers, pen dosim-
                                                     137
               special set-ups with gamma ( Co and  Cs),      eters, survey meters (including superficial
                                            60
               beta ( Sr +  Y,  Tl and  147 Pm), alpha ( 241 Am,   contamination detectors), alarm dosimeters,
                     90
                           90
                              204
               233 U,  238 Pu,  244 Cm, etc.) and low energy X ra-  clinical dosemeters and others. The distribu-
               diations (60 kV).                              tion in terms of levels of calibration is: 13 %
                                                              for radiation therapy, 10% for diagnostic ra-
               Clinical dosimeters (radiotherapy level) can   diology and 77% for radiation protection level.
   62   63   64   65   66   67   68   69   70   71   72