Page 33 - A Practical Guide to Equine Radiography
P. 33

14  A PRACTICAL GUIDE TO EQUINE RADIOGRAPHY



                   It is influenced by the following.                still achieve an acceptable result. CR and DR
      VetBooks.ir  •  Scatter: in the horse, the major factor caus-  systems have a wide latitude, meaning that

                                                                     over- and underexposures can still result in a
                      ing ‘noise’ in an image is scatter radiation,
                                                                     Of course, following the ALARA principle
                      which is in turn influenced by the amount      diagnostic image thanks to post-processing.
                      and radiodensity of the tissues. This is espe-  one should always aim for the lowest expo-
                      cially a problem in the proximal area of the   sure necessary for radiation safety reasons.
                      horse.                                       •  Overexposure on CR and DR systems does
                   •  Collimation: the easiest way to decrease       not result in a black image like on conven-
                      scatter is to collimate as tightly as possible   tional film–screen systems and to assess
                      around the area of interest.                   exposure one needs to be attentive to certain
                   •  Exposure values: discussed below.              numeric values (e.g. in the Fuji CR system
                   •  Detector and post-processing system: some sys-  they are labelled S and L value and they must
                      tems are more sensitive to scatter than others.  be within a certain range). These values are
                   •  Filter: this can be in the form of mechani-    recommended by the manufacturer.
                      cal grids or digital filters. Grids can be used   •  Overexposure can result in an artificial black-
                      to reduce scatter to a certain degree. DR sys-  ness of borders of bones (‘blackout’). This
                      tems often reach the same scatter reduction    may obscure lesions at bone margins and
                      as a grid through inbuilt filters. Grids need to   care should be taken that the soft tissue
                      be carefully chosen for the system used and    envelope is visible on radiographs to ensure
                      require perfect alignment of the beam and an   that the true margins of the bones are visible
                      exact distance between the grid and the X-ray   (Fig. 3.3).
                      machine (required distance is usually shown   •  Underexposure often results in a very noisy
                      on a label on the grid).                       image, which influences our ability to see tra-
                                                                     becular detail and may also obscure lesions;
                                                                     in these cases, an increase in mAs is indicated
                   Exposures                                         (Fig. 3.4).
                   There are three parameters that determine radi-  •  Portable X-ray machines are low-output
                   ation exposure: kVp (peak kilovoltage), mA        generators. The maximum output they
                   (milliamperes) and time (seconds).                are capable of should be displayed on the
                                                                     machine (usually on a little sticker some-
                   •  kVp determines the energy of the X-ray beam    where). Many of these machines will only
                      and its penetration ‘power’. Decreasing the    generate the maximum output for a few
                      kVp settings increases the image contrast and   times in a short time period before they
                      decreases the latitude. A kVp setting under 70   either stop working for a while or image
                      is desirable for good bone radiographs. The    quality will deteriorate because the machine
                      higher the kVp, the lower the contrast and     generates lower outputs.
                      the more scatter is generated.
                   •  mAs is the product of time and mA and
                      determines the number of electrons and       Image distortion and centring
                      hence photons. It influences the ‘blackness’   •  Image distortion is influenced by the angle of
                      of radiographs, but not the contrast. As a rule   the X-ray beam to the object and the angle
                      of thumb, it is a good idea to keep the time to   of the X-ray beam to the plate. Ideally, the
                      under 0.2 seconds to avoid motion artefacts.   X-ray beam should be aligned at 90 degrees
                   •  Exposure latitude is the extent to which a     to the area of interest and the plate to avoid
                      radiograph can be over- or underexposed and    image distortion.











       Equine Radiography.indb   14                                                                              27/11/2018   11:04
   28   29   30   31   32   33   34   35   36   37   38