Page 109 - The Veterinary Laboratory and Field Manual 3rd Edition
P. 109

78  Willy Schauwers

            •  Put the tip firmly on the pipettor.     is leakage. The leakages arise from damage
            •  Press the plunger of the pipettor to the sec-  either to the seals, pistons or tip cones. Often
              ond stop.                                not detectable by the naked eye, leaks lead to
            •  With the pipettor vertical and the tip 1–2 mm   significant volume errors. Air-displacement
              below the surface of the liquid, release the   pipettes must be checked at regular intervals
              plunger smoothly and slowly to aspirate the   and the results must be compared with the
              fluid. Wipe the outside of the tip, do not   error limits. However, a calibration certificate
              touch the open end of the tip.           only reflects the results at the time of testing.
            •  To dispense, touch the tip at an angle against   The time between these calibrations is cru-
              the side of the receiving vessel, press to the   cial, since leaks can occur at any time. While
              first stop and wait a few seconds. Do not   the PLT unit cannot replace regular gravimet-
              press to the second stop (the liquid in the tip   ric testing, daily pipette checks can provide a
              should be discarded).                    safeguard during the periods between calibra-
                                                       tions. Even the smallest leaks are detected!
            Micropipettor tips can be packaged in bulk or in   Process reliability for the pipettes is thus
            tip-boxes (racks, no hand contact with the tip   significantly improved.
            needed). Tips can be sterile, non-sterile and with   •  Every 6 months the micropipettors should
            or without filter.                         be checked for leakage, accuracy and preci-
                                                       sion. If leakage, accuracy and precision are
            carE, uSE and MaIntEnancE                  within limits, the micropipettors can be used
            •  First organize your work station for maxi-  for another 6 months, if the micropipettor
              mum efficiency and minimum fatigue.      fails some parts have to be replaced and it
            •  Make sure the tip is properly mounted and   has to be recalibrated. Some micropipettors
              fits well before you set the volume.     can be re-calibrated; others have to be sent
            •  Adjust the volume.                      back to the manufacturer. Testing micropipet-
            •  Choose the mode of pipetting (reverse or for-  tors requires the use of an analytical balance
              ward mode) adapted to your sample.       (sensitivity 0.0001 g, usually not available in
            •  Eject the used tip and store the pipette in an   a district laboratory) and some specialized
              upright position to avoid damage and cross-  training. See:
              contamination.
            •  Do not leave your pipette lying on the work-  •  ‘SOP-LABORATORY-EQUIP-07 Quality
              bench where it can come into contact with   assurance of mechanical pipettes’
              chemicals or fall off and break. Always store   •  ‘FORM-SOP-LABORATORY-EQUIP-07–01
              your pipette vertically to prevent liquids from   Pipette Calibration Worksheet’
              running inside the shaft of the pipette.  •  ‘FORM-SOP-LABORATORY-EQUIP-07–02
            •  Use the right tip for the pipette, there is a   Gravimetric pipette test worksheet for
              very wide range of tips.                   Pipetman P-1000’.
            •  Do not re-use tips.
                                                     Dispensers
            quaLIty aSSurancE
            •  Air-displacement pipette leakages can be   Bottle top dispensers (Figure 2.43) are widely
              tested using the BRAND PLT unit (Pipette   available for rapid repetitive dispensing of
              Leak Testing unit).  The most frequent cause   reagents. Some dispensers can be autoclaved.
                             7
              of inaccuracy in  air displacement pipettes   Dispensers are available for different volume