The selection, use, maintenance and quality control of laboratory equipment and supplies   73


                Always indicate (on a label) the due date of the   maintenance contracts are stored with the qual-
                half yearly control on the balance.      ity manager. The maintenance company should
                  If the limit value is exceeded, take the fol-  be accredited and should have an ISO 9001 cer-
                lowing steps.                            tificate.
                                                           During the yearly maintenance, the following
                •  Check if the balance is completely level and   tasks are performed:
                  if all weight measurements are properly per-
                  formed.                                •  thorough cleaning of the balance
                •  If there is still a problem, use a different certi-  •  calibrating the balance with certified calibra-
                  fied control weight.                     tion weights
                •  If still a problem, contact the manufacturer of   •  checking the accuracy, precision, linearity and
                  the certified control weights.           eccentric weighing.
                •  If the limit values are exceeded, the pos-
                  sible  consequences  on  earlier  weight   After the maintenance, a calibration report and
                  measurements should be considered.  All   certificate are delivered.
                  measurements after the last successful refer-  The yearly maintenance is registered on
                  ence control weight measurement and before   the form ‘FORM-SOP-LABORATORY-EQUIP-
                  the exceeded measurement should be traced.   01–03-Balance maintenance by maintenance
                  Possible consequences on those laboratory   company’. It is good practice to label each bal-
                  results should be examined and rectified.  ance with the date of the yearly maintenance.
                •  The investigation  of  the  problem and all
                  corrective actions are registered on ‘FORM-
                  SOP-LABORATORY-EQUIP-01–02-Corrective  Water baths
                  actions’.
                                                         Water baths (Figure 2.37) are used to maintain
                yEarLy SErvIcInG                         the temperature of liquids held in laboratory
                If possible, the laboratory should have a con-  containers. The temperature range used for most
                tract for a yearly servicing on each balance. The   purposes is 37°C to 60°C. There is a wide range of



















                                                                    Figure 2.37  Water bath.
                                                                    Photo: Willy Schauwers, Provincial
                                                                    Institute for Hygiene, Antwerp,
                                                                    Belgium.